Objective: To test the safety and efficacy of a bioartificial liver support system in patients with severe acute liver failure.

Summary Background Data: The authors developed a bioartificial liver using porcine hepatocytes. The system was tested in vitro and shown to have differentiated liver functions (cytochrome P450 activity, synthesis of liver-specific proteins, bilirubin synthesis, and conjugation). When tested in vivo in experimental animals with liver failure, it gave substantial metabolic and hemodynamic support.

Methods: Seven patients with severe acute liver failure received a double lumen catheter in the saphenous vein; blood was removed, plasma was separated and perfused through a cartridge containing 4 to 6 x 10(9) porcine hepatocytes, and plasma and blood cells were reconstituted and reinfused. Each treatment lasted 6 to 7 hours.

Results: All patients tolerated the procedure(s) well, with neurologic improvement, decreased intracranial pressure (23.0 +/- 2.3 to 7.8 +/- 1.7 mm Hg; p < 0.005) associated with an increase in cerebral perfusion pressure, decreased plasma ammonia (163.3 +/- 21.3 to 112.2 +/- 9.8 microMoles/L; p < 0.01), and increased encephalopathy index (0.60 +/- 0.17 to 1.24 +/- 0.22; p < 0.03). All patients survived, had a liver transplant, and were discharged from the hospital.

Conclusions: This bioartificial liver is safe and serves as an effective "bridge" to liver transplant in some patients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1243186PMC
http://dx.doi.org/10.1097/00000658-199405000-00012DOI Listing

Publication Analysis

Top Keywords

bioartificial liver
16
severe acute
12
acute liver
12
liver failure
12
liver
10
patients severe
8
porcine hepatocytes
8
liver transplant
8
+/-
6
patients
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!