Purpose: To reduce scatter dose to the contralateral breast during tangential breast irradiation therapy.

Materials And Methods: The authors used a 4-mm-thick commercially available vinyl-coated flexible lead shield that contains a compound of vinyl resin and lead powder of 1-mm-equivalent lead density. Contralateral scatter dose was measured at 5 cm outside the geometric field edge and along the field axis, with and without use of block trays and different wedges.

Results: On the skin, contralateral scatter dose from the medial tangential beam was reduced from about 15% to about 5% of the therapeutic dose delivered to the treated breast. The bulk of scatter dose was absorbed within the first 2 cm of breast tissue.

Conclusion: Reduction in unwanted scatter dose to the contralateral breast is in accord with the philosophy of keeping radiation exposure as low as reasonably achievable and might be of most benefit for young patients.

Download full-text PDF

Source
http://dx.doi.org/10.1148/radiology.191.3.8184078DOI Listing

Publication Analysis

Top Keywords

scatter dose
20
dose contralateral
12
contralateral breast
12
breast tangential
8
tangential breast
8
breast irradiation
8
contralateral scatter
8
dose
7
breast
7
scatter
6

Similar Publications

Purpose: Radiation protection glasses reduce eye lens exposure in Interventional Radiology (IR). However, the protection ratio differs for the lead equivalent content and shape of the lenses. This study aimed to examine factors effective in reducing the lens dose by measuring the protection ratio of scattered radiation due to changes in the surgeon's face orientation, various lead equivalents, and shapes.

View Article and Find Full Text PDF

Purpose: The common approach for organ segmentation in hybrid imaging relies on coregistered CT (CTAC) images. This method, however, presents several limitations in real clinical workflows where mismatch between PET and CT images are very common. Moreover, low-dose CTAC images have poor quality, thus challenging the segmentation task.

View Article and Find Full Text PDF

Position in proton Bragg curve influences DNA damage complexity and survival in head and neck cancer cells.

Clin Transl Radiat Oncol

March 2025

Department of Molecular Genetics, Oncode Institute, Erasmus MC Cancer Institute, Erasmus University Medical Center, Rotterdam, the Netherlands.

Background And Purpose: Understanding the cellular and molecular effect of proton radiation, particularly the increased DNA damage complexity at the distal end of the Bragg curve, is current topic of investigation. This work aims to study clonogenic survival and DNA damage foci kinetics of a head and neck squamous cell carcinoma cell line at various positions along a double passively scattered Bragg curve. Complementary studies are conducted to gain insights into the link between cell survival variations, experimentally yielded foci and the number and complexity of double strand breaks (DSBs).

View Article and Find Full Text PDF

H*10 neutron dosimetry (unlike gamma dosimetry), requires consideration of neutron energy spectra due to the 20× variation of the weight factor over the thermal-to-fast energy range, as well as the neutron radiation field dose rates ranging from cosmic, ~.01 μSv h-1 levels to commonly encountered ~10-200 μSv h-1 in nuclear laboratories/processing plants, and upwards of 104 Sv h-1 in nuclear reactor environments. This paper discusses the outcome of the comparison of spectrum-weighted neutron dosimetry covering thermal-to-fast energy using the novel H*-TMFD spectroscopy-enabled sensor system in comparison with measurements using state-of-the-art neutron dosimetry systems at SRNS-Rotating Spectrometer (ROSPEC), and non-spectroscopic Eberline ASP2E ("Eberline") and Ludlum 42-49B ("Ludlum") survey instrumentation.

View Article and Find Full Text PDF

Analysis of the effect of platelet function and different doses of ticagrelor after flow diverter treatment of intracranial aneurysms.

Neurosurg Rev

January 2025

Neurosurgery Center, Department of Cerebrovascular Surgery, Engineering Technology Research Center of Education Ministry of China on Diagnosis and Treatment of Cerebrovascular Disease, Zhujiang Hospital, Southern Medical University, Guangzhou, China.

Ticagrelor has become the standard drug for the treatment of intracranial aneurysms (IAs) with flow diverters (FDs), but the dosage has not been standardized. The effect of platelet function on clinical and imaging prognosis remains unclear. This study aimed to show the effects of different doses of ticagrelor and platelet aggregation function on the clinical and imaging prognosis after FDs treatment of aneurysms.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!