Many conflicting studies have been performed evaluating the effects of sterilization and storage treatments on the mechanical properties of allograft bone. In the current study, four right and left matched, sterile, deep-frozen, tibial pairs from human donors with an average age of 32 years were tested. One tibia from each pair served as a matched control for the opposite side. Each tibia was cut into four equal segments. One segment of each tibia underwent no treatment; the other three underwent one of the following treatments: irradiation, freeze drying, or ethylene oxide (ETO). Screw pullout tests were performed using four 3.5-mm cortical screws per segment. Sixteen screw pullout tests were performed for each allograft treatment. The freeze-dried specimens required significantly less force for screw pullout. The screw pullout force for the irradiated specimens and the ETO specimens did not significantly differ from their controls. These results indicate that freeze drying affected screw pullout strength. This method of processing should be questioned for structural allografts in which screw fixation is mandatory. The use of irradiation or ETO for sterilization may not have an adverse effect on screw pullout strength.

Download full-text PDF

Source

Publication Analysis

Top Keywords

screw pullout
28
pullout strength
12
sterilization storage
8
storage treatments
8
screw
8
allograft bone
8
freeze drying
8
pullout tests
8
tests performed
8
pullout
7

Similar Publications

Introduction Spinal fusion surgery with pedicle screws is commonly performed to stabilize the spine of osteoporotic patients. However, securing a strong screw fixation in osteoporotic bone presents significant challenges due to the reduced bone density. This study aimed to compare the biomechanical performance in an osteoporotic bone model of pedicle screws inserted using two different techniques, the Jamshidi needle technique and the pedicle probe technique, as well as the influence of tapping on both these techniques.

View Article and Find Full Text PDF

A curved compliant spinal bone anchor to enhance fixation strength.

PLoS One

December 2024

Bio-Inspired Technology Group, Faculty of Mechanical Engineering, Department of BioMechanical Engineering, Delft University of Technology, Delft, The Netherlands.

Pedicle screws have long been established as the gold standard for spinal bone fixation. However, their fixation strength can be compromised in cases of low bone density, particularly in osteoporotic bone, due to the reliance on a micro-shape lock between the screw thread and the surrounding bone. To address this challenge, we propose augmenting conventional pedicles screws with a curved compliant anchor.

View Article and Find Full Text PDF

Characterizing orthodontic mini-screws in the hard palate of pigs: An experimental and finite-element study.

Heliyon

February 2024

Department of Life, Health and Environmental Sciences, Postgraduate School of Orthodontics, Università degli Studi dell'Aquila, Via Piazzale Salvatore Tommasi 1, Abruzzo, L'Aquila 67100, Italy.

Objective: This paper analyses the mechanical response of mini-implants under pull-out, push-in, and shear forces.

Materials And Methods: The authors have devised a specialized testing apparatus, using a universal testing machine, for the mechanical characterisation of orthodontic mini-implants (OMI) installed in pig hard palate. The experimental investigation encompasses seven screw types, each subjected to pull-out, push-in, and shear forces.

View Article and Find Full Text PDF

Triangular pelvic fixation in the management of adult spinal deformities: technical note and preliminary results.

Orthop Traumatol Surg Res

December 2024

Aix-Marseille Université, APHM, CNRS, ISM, CHU Timone, Service de Chirurgie Orthopédique et Vertébrale, 264 rue Saint-Pierre, 13005 Marseille, France. Electronic address:

Article Synopsis
  • The choice of the lower vertebra for spinal fixation in adult spinal deformity, particularly around the lumbosacral junction, is debated, with options including stopping fixation at S1 or using lower points like S2AI or iliac screws to prevent implant pullout.
  • A new technique for posterior fixation combines S1 screws, iliac screws, and sacro-iliac implants, inspired by triangular pelvic fixation methods used for trauma.
  • This study outlines the operative technique using fluoroscopic guidance and shares preliminary results from treating 15 patients, labeled as level IV evidence.
View Article and Find Full Text PDF

An OLIF Cage Integrated with a Low-Profile Plate and Cross Screws Could Reduce the Risk of Postoperative Complications Biomechanically.

Ann Biomed Eng

December 2024

Luzhou Key Laboratory of Orthopedic Disorders, Department of Orthopedics, The Affiliated Traditional Chinese Medicine Hospital, Southwest Medical University, No. 182, Chunhui Road, Longmatan District, Luzhou, 646000, Sichuan, P.R. China.

Article Synopsis
  • Stand-alone oblique lumbar interbody fusion (OLIF) has stability issues post-surgery, prompting the use of additional fixation devices (AFDs) like anterior lateral single rod (ALSR) fixation, which come with potential complications.
  • Researchers developed a new OLIF cage with a low-profile plate and cross screw (LPCS) to enhance stability without requiring changes in body position or extra incisions during surgery.
  • Testing showed that the LPCS OLIF cage significantly outperformed ALSR fixation in terms of bending and pullout strength, with LPCS models demonstrating a much higher resistance to failure, particularly beneficial for patients with osteoporotic bone density.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!