Thermal effects on meniscus and bone tissue after application of 314 boreholes using five different infrared (IR) lasers: Nd:YAG, Tm:YAG, Ho:YAG, Er:YAG, Cr,Er:YSGG (application energy 200 mJ, 400 mJ, 600 mJ, 800 mJ, 1000 mJ; repetition rate 2 Hz, 5 Hz; medium air, water rinse) were analyzed. The experimental set-up comprised for the beam guiding a focussing lens (f = 100 mm) or a flexible fiber (Ho:YAG). Damaged tissue was investigated macroscopically, histologically, and by scanning electron microscopy. Application in air caused carbonisation in all cases. Application in water showed thermal brown discoloring using Tm:YAG, Ho:YAG laser on meniscus tissue. The Nd:YAG did not ablate. The Er:YAG laser showed macroscopically precise boreholes without any discoloring of the adjacent tissue as well in meniscus as in bone. Cr:ErYSGG laser results were comparable with the results using an Er:YAG laser although ablation on bone tissue created higher thermal effects. For the aim of developing minimal invasive operating techniques the Er:YAG laser showed best results.