Development of a patient-dedicated, on-demand, blood gas monitor.

Am J Respir Crit Care Med

Department of Medicine, Veterans Affairs Medical Center, Long Beach, CA 90822.

Published: April 1994

AI Article Synopsis

  • A new bedside monitor (CDI 2000) has been developed for real-time blood gas measurements, allowing blood to be drawn and returned with minimal loss and infection risk.
  • In vitro tests using bovine blood showed the monitor had very small measurement biases, indicating high accuracy compared to traditional methods.
  • Trials in normal volunteers also demonstrated the monitor's performance is on par with established blood gas analyzers, suggesting it could be a reliable alternative in clinical settings.

Article Abstract

A new monitor (CDI 2000) that brings blood gas measurements to the patient's bedside has been developed. To measure blood gases, blood is drawn into the patient's arterial pressure-monitoring line past in-line fluorescent-based sensors. After measurement, the blood is returned to the patient, avoiding blood loss and delays in sample turnaround and reducing the risk of infection to both patient and operator. We assessed this system's performance in vitro with tonometered bovine blood. Bias (mean difference between monitor and tonometered gas or measured pH values) +/- the standard deviation (SD) were 0.01 +/- 0.02 at pH = 7.39; 0.0 +/- 0.7 mm Hg at Pco2 = 39 mm Hg; and 2.4 +/- 3.2 mm Hg at a Po2 = 100 mm Hg (n = 54). Changes in hematocrit, blood temperature, or serum sodium concentration did not have clinically significant effects on system performance. Studies in normal volunteers, in whom large changes in blood gases were induced, showed a bias (mean difference between monitor and IL 1306 values) +/- SD of 0.00 +/- 0.02 for pH, -0.4 +/- 2.0 mm Hg for Pco2, and -3.6 +/- 7.7 mm Hg for Po2 (n = 69). We conclude from the present study that the performance of this system is comparable to that of conventional blood gas analyzers.

Download full-text PDF

Source
http://dx.doi.org/10.1164/ajrccm.149.4.8143046DOI Listing

Publication Analysis

Top Keywords

blood gas
12
blood
10
blood gases
8
bias difference
8
difference monitor
8
+/-
8
values +/-
8
+/- 002
8
+/- pco2
8
+/- po2
8

Similar Publications

Histopathology of the small airways: Similarities and differences between ageing and COPD.

Pulmonology

December 2025

Division of Immunology, Immunity to Infection and Respiratory Medicine, School of Biological Sciences, Faculty of Biology, Medicine and Health, University of Manchester and Manchester University NHS Foundation Trust, Manchester, UK.

Age-related lung function decline is associated with small airway closure and gas trapping. The mechanisms which cause these changes are not fully understood. It has been suggested that COPD is caused by accelerated ageing.

View Article and Find Full Text PDF

European central hypoventilation syndrome consortium description of congenital central hypoventilation syndrome neonatal onset.

Eur J Pediatr

January 2025

Service de Physiologie Pédiatrique-Centre du Sommeil-CRMR Hypoventilations Alvéolaires Rares, INSERM NeuroDiderot, Université Paris-Cité, AP-HP, Hôpital Robert Debré, Paris, France.

Unlabelled: It is known that in most cases of congenital central hypoventilation syndrome (CCHS), apnoeas and hypoventilation occur at birth. Nevertheless, a detailed description of initial symptoms, including pregnancy events and diagnostic tests performed, is warranted in infants with neonatal onset of CCHS, that is, in the first month of life. The European Central Hypoventilation Syndrome Consortium created an online patient registry from which 97 infants (44 females) with CCHS of neonatal onset and PHOX2B mutation from 10 countries were selected.

View Article and Find Full Text PDF

Accuracy of two-compartment modelling of gas exchange with ventilation-perfusion mismatch in inhalational anesthesia.

Anesthesiology

January 2025

Department of Critical Care, Melbourne Medicine School, Faculty of Medicine, Dentistry and Health Sciences, University of Melbourne, Victoria, Australia.

Background: Multi-compartment computer models of heterogeneity in alveolar ventilation-perfusion ratios (VA/Q scatter) across the lung explain the significant alveolar-arterial (A-a) partial pressure gradients and associated alveolar dead-space fractions (VDA/VA) seen in anesthetized patients for both carbon dioxide and for anesthetic gases of different blood solubilities. However, the accuracy of a simpler two-compartment model of VA/Q scatter to do this has not been tested or compared to calculations from the traditional Riley model with "ideal", unventilated (shunt) and unperfused (deadspace) compartments.

Methods: Measurements of gas partial pressures in inspired and expired gas and arterial and mixed venous blood from 29 patients undergoing inhalational general anesthesia for cardiac surgery was used to compare the accuracy of two simple models of VA/Q scatter and lung gas exchange in predicting measured alveolar and arterial partial pressure differences, and associated alveolar dead-space calculations for the modern anesthetic gases isoflurane, sevoflurane and desflurane.

View Article and Find Full Text PDF

Objectives: The aim of this study was to evaluate the factors associated with non-invasive mechanical ventilation (NIMV) failure in acute cardiogenic pulmonary edema (ACPE) diagnosed in the emergency department.

Methods: This study was prospectively conducted at the Ege University Faculty of Medicine ED between February 19, 2021 and December 01, 2021. Patients who received NIMV with ACPE were included.

View Article and Find Full Text PDF

Fast is fine, but accuracy is everything: making intraoperative transfusion decisions using point-of-care testing.

Br J Anaesth

February 2025

Transfusion Research Unit, Department of Public Health and Preventative Medicine, Monash University, Melbourne, VIC, Australia; Department of Clinical Haematology, Monash Health, Clayton, VIC, Australia.

Accurate and timely diagnostic information is a vital adjunct to clinical assessment to inform therapeutic decision-making, including decisions to transfuse, or not transfuse, blood components. A prospective cohort study of diagnostic point-of-care (POC) haemoglobin measurements on arterial or central venous samples from adults undergoing major noncardiac surgery compared three widely used devices, HemoCue®, i-STAT™, and the Rad-67™ pulse CO-Oxymeter® finger sensor device, against standard laboratory haemoglobin measurements, but importantly not against a blood gas analyser. The study focused on haemoglobin results below 100 g L to establish the utility of these devices to guide red cell transfusion decisions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!