AI Article Synopsis

  • The study assessed the effectiveness of extracorporeal high-intensity focused ultrasound (HIFU) on both normal and tumor-bearing rabbit livers using varying intensities and exposure times.
  • In the experiments, rabbits were divided into groups to observe the impact of different SPTP intensities and their relationships with lesion volume and healing.
  • Results showed that lower intensity (500 W cm-2) led to better tumor destruction compared to higher intensity, while increased exposure time consistently resulted in larger lesions, with some adverse effects observed at extremely high intensities.

Article Abstract

Treatment parameters of extracorporeal high intensity focused ultrasound (HIFU) were analysed in normal and tumor-bearing rabbit liver. HIFU was generated with a 1 MHz transducer and energy was provided by a 7.5 kW power amplifier. In vivo experiments were conducted on 74 New Zealand rabbits. Normal rabbits and rabbits bearing an intrahepatic VX2 tumor were used. In group 1, spatial peak temporal peak (SPTP) intensities ranging from 1470 to 5500 W cm-2 and exposure times from 0.5 to 5 s were tested at a constant depth in the liver; in group 2, the output power was adjusted as a function of the target depth in order to keep constant the focal in situ intensity in the liver; in group 3 (liver tumors), the focal in situ intensity was 1365 W cm-2 in eight rabbits and 500 W cm-2 in nine. In groups 1, 2 and 3, rabbits were sacrificed 48 h after the treatment. Groups 4 and 5 were designated for analysis of the lesion in the normal liver 4 weeks after treatment at 1000 W cm-2 and 3000 W cm-2 SPTP intensities, respectively. In normal rabbits, the lesion volume increased with exposure time at constant intensity; there was a negative correlation between intensity and exposure time (group 1). When the output power was adjusted as a function of the path length, the lesion size was nearly constant (group 2). In VX2 rabbits, tumor destruction rates were significantly higher in rabbits treated at 500 W cm-2 than in rabbits treated at 1365 W cm-2 (p < 0.05; group 3). As in the normal liver, the lesion volume increased with the exposure time at constant intensity. HIFU lesions treated at 1000 w cm-2 (SPTP) healed as thin fibrous scars, and no severe complication occurred (group 4); at 3000 W cm-2 (SPTP), scars were larger and perforation of a neighbouring organ was seen in 7 of 11 rabbits (group 5).

Download full-text PDF

Source
http://dx.doi.org/10.1016/0301-5629(93)90096-7DOI Listing

Publication Analysis

Top Keywords

cm-2 sptp
12
exposure time
12
rabbits
10
cm-2
9
normal tumor-bearing
8
high intensity
8
intensity focused
8
focused ultrasound
8
normal rabbits
8
group
8

Similar Publications

Despite the use of therapeutic ultrasound in the treatment of soft tissue pathologies, there remains some controversy regarding its efficacy. In order to develop new treatment protocols, it is a common practice to carry out in vitro studies in cell cultures before conducting animal tests. The lack of reproducibility of the experimental results observed in the literature concerning in vitro experiments motivated us to establish a methodology for characterizing the acoustic field in culture plate wells.

View Article and Find Full Text PDF

Improving the Function and Engraftment of Transplanted Pancreatic Islets Using Pulsed Focused Ultrasound Therapy.

Sci Rep

September 2019

Interventional Regenerative Medicine and Imaging Laboratory, Stanford University School of Medicine, Department of Radiology, Palo Alto, California, 94304, USA.

This study demonstrates that pulsed focused ultrasound (pFUS) therapy can non-invasively enhance the function and engraftment of pancreatic islets following transplantation. In vitro, we show that islets treated with pFUS at low (peak negative pressure (PNP): 106kPa, spatial peak temporal peak intensity (I): 0.71 W/cm), medium (PNP: 150kPa, I: 1.

View Article and Find Full Text PDF

Background: Occlusion of blood vessels using high-intensity focused ultrasound (HIFU) is a potential treatment for arteriovenous malformations and other neurovascular disorders. However, attempting HIFU-induced vessel occlusion can also cause vessel rupture, resulting in hemorrhage. Possible rupture mechanisms include mechanical effects of acoustic cavitation and heating of the vessel wall.

View Article and Find Full Text PDF

This paper focuses on the development of a finite-element model and subsequent stationary analysis performed to optimize individual flexural piezoelectric elements for operation in the frequency range of 20-100kHz. These elements form the basic building blocks of a viable, un-tethered, and portable ultrasound applicator that can produce intensities on the order of 100mW/cm(2) spatial-peak temporal-peak (I(SPTP)) with minimum (on the order of 15V) excitation voltage. The ultrasound applicator can be constructed with different numbers of individual transducer elements and different geometries such that its footprint or active area is adjustable.

View Article and Find Full Text PDF
Article Synopsis
  • This paper focuses on optimizing lightweight, battery-operated ultrasound transducers for biomedical use, targeting frequencies of 20-100kHz.
  • The aim is to create a wearable ultrasound applicator that operates on low excitation voltages (10-25V) while delivering sufficient acoustic pressure (approximately 55kPa or 100mW/cm²) for therapeutic applications like chronic wound management and drug delivery.
  • Early prototypes were designed using complex impedance analysis and are powered by portable lithium polymer batteries, enabling features like adjustable duty cycles and continuous or pulsed operation.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!