Evaluation of a turbidimetric procedure for apolipoproteins A1 and B on the Ciba Corning 550 EXPRESS.

Clin Biochem

Department of Laboratory Medicine, Hamilton General Division, Hamilton Civic Hospitals, Ontario, Canada.

Published: December 1993

We evaluated a turbidimetric method for the estimation of apo A1 and apo B on the Ciba Corning 550 EXPRESS using Ciba Corning reagents. Interference due to bilirubin, hemoglobin, lipemia, triglycerides, and uremia was minimal, with apolipoproteins (apo) A1 and B results usually within +/- 4% of expected values. Within-run and day-to-day imprecision (coefficients of variation) ranged from 1.96 to 3.60% and 2.63 to 3.39% for apo A1 and 1.02 to 1.74% and 2.08 to 3.66% for apo B, respectively. Accuracy was determined by participation in the IFCC apolipoprotein standardization project in which results obtained on 50 patient samples were compared to those obtained by the reference laboratory. Apo A1 and apo B showed an average bias of +3.7% and +2.0% and correlation coefficients of 0.986 and 0.977, respectively. Results were also compared to those obtained on the Behring Turbitime system and showed a bias of +7.5% and -8.8% for apo A1 and apo B, respectively. The Ciba Corning automated method was rapid and gave good accuracy, precision, linearity, and parallelism and was relatively unaffected by raised triglyceride values.

Download full-text PDF

Source
http://dx.doi.org/10.1016/0009-9120(93)80012-jDOI Listing

Publication Analysis

Top Keywords

ciba corning
16
apo apo
12
apo
9
corning 550
8
550 express
8
apo ciba
8
evaluation turbidimetric
4
turbidimetric procedure
4
procedure apolipoproteins
4
ciba
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!