  • The study focuses on measuring the distance between a gynecological afterloading applicator used in HDR-Ir-therapy and the rectum mucosa.
  • A Co-60 source was utilized to assess maximum doses at the rectum by withdrawing the source in 1 cm increments, ensuring doses do not exceed 3.0 Gy.
  • In 91 patients, distances from the rectum mucosa to the applicator ranged from 5 to 18 mm, with an average of 14 mm, confirming the method's effectiveness in preventing rectum overdose.

Article Abstract

Purpose: Assessment of the distance from a gynecological afterloading applicator to the rectum mucosa.

Methods: The appropriate afterloading applicator for HDR-Ir-therapy is used for that purpose. A Co-60-source (370 MBq) with a guiding wire is positioned at the tip of the applicator and then withdrawn in steps of 1 cm. For each of these points the maximum dose at the rectum mucosa is measured with an ionisation chamber. From a reference curve it is then possible to determine the distance from the rectum mucosa to the applicator and to mark it in the calculated isodose plan. The actual daily dose to the rectum mucosa is planned not to exceed 3.0 Gy in the maximum.

Results: Up to now the measurements were performed in 91 patients. The minimal distances of the rectum mucosa from the applicator were found to range from 5 to 18 mm with an average value of 14 mm.

Conclusions: The comparison of the measurement with the readings of the rectum probe during each session showed, that the method is suitable to avoid an overdose at the rectum mucosa.

Download full-text PDF

Source

Publication Analysis

Top Keywords

rectum mucosa
20
afterloading applicator
12
dose rectum
8
mucosa applicator
8
rectum
7
mucosa
6
applicator
6
[measurement distance
4
distance rectal
4
rectal mucosa
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!