Recommendations for measurement of tray and wedge factors for high energy photons.

Med Phys

Department of Radiation Oncology, E. W. Sparrow Hospital, Lansing, Michigan 48909.

Published: April 1994

Transmission factors for beam modifers such as wedges and blocking trays in dosimetry calculations are very common, but various methods have been used in measuring these factors. A systematic study to measure these factors as a function of field size and point of measurement was carried out. The beam modifiers were a blocking tray and the standard wedges supplied with a cobalt-60 unit and a medical electron accelerator (6- and 18-MV x rays). Several points of measurement were selected, using several depths: dmax, TG-21 calibration depth, 10 cm (wedge angle defining depth, ICRU), 15 cm and in-air (miniphantom), and various field sizes. The results show a small field size dependence in these transmission factors with more dependence found in the point of measurement.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.597392DOI Listing

Publication Analysis

Top Keywords

transmission factors
8
field size
8
point measurement
8
factors
5
recommendations measurement
4
measurement tray
4
tray wedge
4
wedge factors
4
factors high
4
high energy
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!