AI Article Synopsis

Article Abstract

An automated technique involving switching valves and a filter assembly has been developed and evaluated for the on-line precipitation of proteins and peptides from plasma samples. In the set-up, the proteins were precipitated on-line by injecting the plasma sample into a stream of organic precipitating agent. The precipitates so formed are filtered on-line by a set of filter assemblies consisting of ordinary in-line HPLC solvent filters. Evaluation of the technique was performed using ibuprofen and a mixture of three estrogens, estradiol, equilin and estrone, spiked in dog plasma. The coefficients of variation (C.V.) for system suitability parameters were below 10%. Absolute recovery of ibuprofen in plasma ranged from 80% for 100 micrograms/ml to 114% for 5 micrograms/ml spiked concentrations, respectively. Resolution for equilin and estrone, two closely eluting peaks, was 1.79 (C.V. = 5.8%, n = 7). The switching-valve--filter assembly had no significant effect on the efficiency of the HPLC system.

Download full-text PDF

Source
http://dx.doi.org/10.1016/0378-4347(93)e0372-wDOI Listing

Publication Analysis

Top Keywords

equilin estrone
8
plasma
5
switching-valve--filter technique
4
technique direct
4
direct injection
4
injection analysis
4
analysis drugs
4
drugs plasma
4
plasma high-performance
4
high-performance liquid
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!