The Pulsatile catheter (PUCA) pump consists of a single port membrane pump connected to an indwelling valved catheter. This so-called transarterial blood pump was originally designed to be introduced through a superficial artery into the left ventricular cavity to pump blood from the left ventricle into the ascending aorta. By introducing the catheter directly into the thoracic aorta or the pulmonary artery, the possibility is created of applying large-diameter catheter PUCA pumps as left, right, or biventricular assist devices (LVAD, RVAD, or BIVAD) without damaging any of the structures of the heart. The pump performance of an 8 mm PUCA pump prototype (internal diameter catheter, 8 mm; catheter length, 40 cm; stroke volume, 80 ml) was studied in a mock circulation to investigate the influence of pulsatile intraventricular pumping on ventricular pressure patterns. The pumping mode of the PUCA pump was changed from approximately 1:1 ([n + 1]:n) to 1:2 ([1/2n + 1]:n) and 1:3 ([1/3n + 1]:n) in relation to the frequency of a ventricle-simulating membrane pump. Apart from the pumping mode, timing of the PUCA pump driving system (ejection phase) seems to be crucial in obtaining optimal unloading of the ventricle.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1525-1594.1994.tb03366.xDOI Listing

Publication Analysis

Top Keywords

puca pump
16
pump
9
influence pulsatile
8
pulsatile intraventricular
8
intraventricular pumping
8
pumping ventricular
8
ventricular pressure
8
pressure patterns
8
catheter puca
8
membrane pump
8

Similar Publications

Mechanical ventilation and thoracic artificial lung assistance during mechanical circulatory support with PUCA pump: in silico study.

Comput Methods Programs Biomed

February 2014

Department of Cardiovascular, Respiratory, Nephrologic, Anaesthesiologic and Geriatric Sciences, University "Sapienza" of Rome, Italy; National Institute of Cardiovascular Research, Bologna, Italy.

Patients assisted with left ventricular assist device (LVAD) may require prolonged mechanical ventilatory assistance secondary to postoperative respiratory failure. The goal of this work is the study of the interdependent effects LVAD like pulsatile catheter (PUCA) pump and mechanical ventilatory support or thoracic artificial lung (TAL), by the hemodynamic point of view, using a numerical simulator of the human cardiovascular system. In the simulator, different circulatory sections are described using lumped parameter models.

View Article and Find Full Text PDF

The pulsatile catheter pump (PUCA pump) is a left ventricular assist device that provides additional flow to the left ventricle. It is usually run in order to ensure a counterpulsation effect, as in the case of the intra-aortic balloon pump (IABP). Because of this similarity, a comparison between the PUCA pump and the IABP was conducted from both the hemodynamic and energetic points of view.

View Article and Find Full Text PDF

This study was aimed to mimic clinical heart failure (HF) conditions and to assess the effect of pulsatile catheter (PUCA) pump support on hemodynamics and tissue perfusion in a sheep model of acute HF. In 14 sheep, HF was induced by partial occluding the middle left circumflex coronary artery combined with pacemaker-induced tachycardia. PUCA pump was then activated to support the HF for 3 h.

View Article and Find Full Text PDF

The PUCA II pump is a minimally invasive intra-arterial left ventricular assist device that can be used as an alternative for the intra-aortic balloon pump (IABP). In this study, we assessed the cardiac unloading and organ perfusion capacities of both PUCA II and IABP in an in vitro set up, consisting of a heart simulator and a silicone arterial tree, mimicking anatomical geometry and flow distribution. The IABP was positioned in the descending aorta, while the PUCA II was tested both in 'trans-aortic' and 'abdominal' positions.

View Article and Find Full Text PDF

In vitro evaluation of the PUCA II intra-arterial LVAD.

Int J Artif Organs

August 2003

Hydraulics Laboratory, Institute Biomedical Technology, Ghent University, Ghent, Belgium.

The "pulsatile catheter" (PUCA) pump is a minimally invasive intra-arterial left ventricular assist device intended for acute support of critically ill heart failure patients. To assess the hydrodynamic performance of the PUCA II, driven by an Arrow AutoCat IABP driver, we used a (static) mock circulatory system in which the PUCA II was tested at different loading conditions. The PUCA II was subsequently introduced in a (dynamic) cardiovascular simulator (CVS) to mimic actual in vivo operating conditions, with different heart rates and 2 levels of left ventricular (LV) contractility.