A method for determining the selenium content of infant formulas is proposed. It includes wet digestion with nitric acid and hydrogen peroxide in medium pressure teflon bombs in a microwave oven and determination by graphite furnace atomic absorption spectrometry (GFAAS). The absence of interferences is checked. Values obtained for the limit of detection (19.4 ng/g), precision (RSD = 2.2%) and accuracy by analysis of a reference material show that the method is reliable.