Attenuation correction in PET using single photon transmission measurement.

Med Phys

Department of Physics, McMaster University, Hamilton, Ontario, Canada.

Published: June 1994

The use of single photon transmission measurement with a rotating rod source has been evaluated to measure the attenuation correction factors in positron emission tomography (PET). The singles projections are resampled into the coincidence geometry using the detector positions and the rod source location. A nonparalyzable dead time correction algorithm was developed for the block detectors used in the McMaster PET scanner. This enables accurate attenuation correction factors (ACFs) to be computed using a wide range of source strengths for transmission scanning. Transaxial resolution is approximately 6 mm, which is comparable to emission scanning performance. Axial resolution is about 25 mm, with only crude source collimation. ACFs are underestimated by as much as 10% due to increased cross-plane scatter, compared to coincidence transmission accuracy. The response of the correction factors to object density is within 15%, when comparing singles transmission measurement to current coincidence transmission measurement. The major advantage of using singles transmission measurement is a dramatically increased count rate. A factor of 7 increase in count rate over coincidence scanning is possible with a 2-mCi transmission rod source. Uniformity of 2% in the transmission images is possible with this source strength and a 2-min acquisition. There are no randoms counted in singles transmission scans, which makes the measured count rate vary linearly with source activity. Singles detector dead time losses are approximately 6% in the detectors opposite a 2-mCi rod source.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.597394DOI Listing

Publication Analysis

Top Keywords

transmission measurement
20
rod source
16
attenuation correction
12
correction factors
12
singles transmission
12
count rate
12
transmission
10
single photon
8
photon transmission
8
source
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!