An empirical calculation method for high-energy beam peripheral dose estimation is described. The peripheral dose has been measured for a Siemens Mevatron MX6700 (6 MV) and a Siemens Mevatron KDS (6 and 18 MV) linear accelerators. The dose distribution is parameterized for each beam energy as a function of depth, distance from the edge of the field, and field size. A simple algorithm has been developed for dose calculation up to 100 cm from the field central axes. Predictions by this algorithm are compared with measurements in an Alderson phantom.