Purpose: In an investigational procedure, excimer laser photorefractive keratectomy for severe myopia was performed at three clinical trial centers to determine the effectiveness of the multiple zone technique.

Methods: A VisX Model Twenty/Twenty excimer laser (VisX, Santa Clara, California) was used to perform photorefractive keratectomy on 14 severely myopic eyes (-10.37 to -24.5 diopters) of 12 patients by using a multiple zone technique. Postoperative follow-up ranged from six months to two years; retreatments were performed on four patients, with a follow-up of at least nine months.

Results: At six months postoperatively, before retreatment, three of the 14 eyes were within 2 diopters and seven of the 14 eyes were within 4 diopters of attempted correction. Regression of effect to more severe myopia was worse in five eyes treated with nitrogen gas blowing. Retreatments also demonstrated considerable myopic regression. Three patients had loss of two or more lines of best-corrected visual acuity, and these patients also had moderate or severe levels of haze.

Conclusion: Excimer laser photorefractive keratectomy for severe myopia using a multiple zone technique is associated with considerable regression, haze, and loss of best-corrected visual actuity, especially when performed in association with nitrogen gas blowing.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0002-9394(14)71166-3DOI Listing

Publication Analysis

Top Keywords

excimer laser
16
photorefractive keratectomy
16
multiple zone
16
severe myopia
16
laser photorefractive
12
zone technique
12
keratectomy severe
8
eyes diopters
8
nitrogen gas
8
gas blowing
8

Similar Publications

Boosting the Pixel Acquisition Rate of Elemental Mapping with Laser Ablation-ICP-Mass Spectrometry to 1000 Hz.

Anal Chem

March 2025

Department of Chemistry, Atomic and Mass Spectrometry - A&MS Research Group, Ghent University, Campus Sterre, Krijgslaan 281-S12, Ghent 9000, Belgium.

Novel low-dispersion ablation cell designs and highly efficient aerosol transport systems have enabled fast elemental mapping using laser ablation-ICP-mass spectrometry (LA-ICP-MS) at high spatial resolution and its application in various research fields. Nowadays, the fastest low-dispersion setups enable narrow single pulse responses (SPR, duration of the transient signal observed upon a single laser shot), which enhance the signal-to-noise ratio and boost the pixel acquisition rate attainable in elemental mapping applications. In this work, the analytical performance of a nanosecond 193 nm ArF* excimer-based kHz laser in combination with a low-dispersion tube-type ablation cell, coupled to an ICP-mass spectrometer equipped with a time-of-flight (ToF) analyzer, was evaluated.

View Article and Find Full Text PDF

Effects of effective optical zone and decentration on visual quality after smile for different astigmatism types.

BMC Ophthalmol

March 2025

State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Guangdong Provincial Key Laboratory of Ophthalmology and Visual Science, Guangdong Provincial Clinical Research Center for Ocular Diseases, Sun Yat-sen University, Guangzhou, 510060, China.

Purpose: To compare the refractive outcomes and visual quality among different types of astigmatism following SMILE and evaluate effective optical zone (EOZ) features, decentration and their potential effects on visual quality.

Methods: This study included 101 left eyes of 101 patients who underwent SMILE. Patients were grouped according to astigmatism types (with-the-rule [WTR], against-the-rule [ATR] and oblique astigmatism) and decentered displacement (major axis > minor axis and major axis < minor axis).

View Article and Find Full Text PDF

ObjectivesProximal to mid-term effects of contrast excimer laser atherectomy (ELA) + drug-coated balloon (DCB) versus PTA + drug-coated balloon (DCB) in the treatment of new lesions in the femoral popliteal segment of patients with lower extremity arteriosclerosis and occlusion.MethodsThe clinical data of 105 patients with CT-confirmed femoropopliteal segment lesions (Rutherford grades 3-6) were retrospectively analyzed. According to the computerized randomization method, 58 patients (44 males, mean 72.

View Article and Find Full Text PDF

Background: Vitiligo, a prevalent acquired depigmentation condition, involves a progressive melanocyte loss, posing significant treatment challenges. Excimer light has been employed in vitiligo treatment, while trichloroacetic acid (TCA) has been noted to induce chemical trauma, leading to perifollicular pigmentation and perilesional repigmentation in vitiligo. This study aims to assess and compare the efficacy and safety of using 308 nm monochromatic excimer light alone versus combining it with 50% TCA in treating vitiligo.

View Article and Find Full Text PDF

Purpose: To evaluate the repeatability of angle kappa measurements obtained from the Pentacam (Oculus Optikgeräte GmbH), Sirius (Costruzione Strumenti Oftalmici), and Keratron Scout (BVI), and to assess the agreement with those obtained from the Amaris 750S (SCHWIND eye-tech-solutions).

Methods: This study included 110 patients scheduled to undergo transepithelial photorefractive keratectomy or femtosecond laser-assisted laser in situ keratomileusis. Before surgery, each patient underwent angle kappa measurement using three instruments: Pentacam, Sirius, and Keratron.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!