Three automated immunoassays for digoxin in serum were evaluated--Abbott TDxII, Baxter Stratus, and Behring OPUS. The accuracy and precision of the assays were assessed by weighed-in controls and an external quality control program. Coefficients of variation of all methods in serum were < or = 10% at weighed-in concentrations of digoxin of 1 and 2.5 micrograms/L. Accuracy relative to weighed-in concentrations of 1 and 2.5 micrograms/L ranged from 98 to 126% for all methods. Comparative results from patient samples showed little difference between the TDxII and Stratus and a greater difference observed between the TDxII and OPUS assays. The detection of digoxin-free samples containing digoxin-like immunoreactive substances (DLIS) in neonatal cord blood, pregnant patients, and liver and renal recipients by each assay was then assessed. The TDxII exhibited the highest incidence of DLIS. This is evident in neonatal cord blood in which 40.4% of samples tested positive. In comparison, the extent of DLIS detected by Stratus was less and OPUS exhibited no DLIS in any of the groups studied. A case study of a patient treated with anti-digoxin Fab fragments (Digibind) also was included for analysis by each method. Fourteen hours after Digibind administration, the TDxII registered a digoxin concentration of 49.5 micrograms/L compared with 3.73, 1.80, and 2.49 micrograms/L for Stratus, OPUS, and ultrafiltered TDxII methods, respectively. The results indicate that to determine the concentration of digoxin after the administration of Digiband, the OPUS or fluorescence polarization immunoassay (FPIA)-ultrafiltered samples by TDxII are the assays of choice.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/00007691-199502000-00011 | DOI Listing |
Clin Chem
February 2004
Laboratorio Analisi Chimico Cliniche 1, Azienda Ospedaliera 'Spedali Civili', 25125 Brescia, Italy.
Background: The European Society of Cardiology/American College of Cardiology Committee for the Redefinition of Myocardial Infarction (MI) has recommended that an increased cardiac troponin should be defined as a measurement above the 99th percentile value of the reference group. A total imprecision (CV) at the decision limit of =10% is recommended. However, peer-reviewed data on assay imprecision are lacking.
View Article and Find Full Text PDFClin Chem
October 2000
Dipartimento di Medicina di Laboratorio, Università-Ospedale, 35100 Padova, Italy.
Background: Lacking assay standardization, different myoglobin methods may produce results that differ significantly.
Methods: A multicenter study was carried out to compare the analytical performance of five commercially available assays for myoglobin measurement. Linearity, imprecision, interferences, and method comparison were studied according to NCCLS guidelines, whereas reference values were determined following IFCC recommendations.
Clin Chim Acta
March 2000
Department of Medicine, University Hospital Charité/Campus Virchow-Klinikum, Humboldt University, Berlin, Germany.
It was suggested recently that cardiac troponins are released as T-I-C complexes and then further degraded to T and I-C. It is not known whether the various affinity to the T-I-C and I-C complex of different troponin I test systems influence the diagnostic and prognostic value of the test results in clinical practice. We studied 162 patients (61.
View Article and Find Full Text PDFClin Chem
July 1999
Spectral Diagnostics Inc., 135-2 The West Mall, Toronto, Ontario, Canada M9C 1C2.
Background: Up to a 20-fold variation in serum cardiac troponin I (cTnI) concentration may be observed for a given patient sample with different analytical methods. Because more limited variation is seen for control materials and for purified cTnI, we explored the possibility that cTnI was present in altered forms in serum.
Methods: We used four recombinantly engineered cTnI fragments to study the regions of cTnI recognized by the Stratus(R), Opus(R), and ACCESS(R) immunoassays.
Clin Chem
June 1999
Department of Clinical Biochemistry, St. Bartholomew's and the Royal London School of Medicine and Dentistry, Turner Street, London E1 2AD, UK. P.O. Box 6101, Newark, DE 19714-610.
Background: Available assays for cardiac troponin I (cTnI) yield numerically different results. The aim of this study was to compare patient values obtained from four cTnI immunoassays.
Methods: We studied the Stratus(R) II assay, the Opus(R) II assay, the Access(R) assay, and a research-only cTnI heterogeneous immunoassay that uses the Dade Behring aca(R) plus immunoassay system equipped with two new noncommercial monoclonal antibodies.