Background: At fibreoptic bronchoscopy the potential exists for contamination of bronchoscopes and microbiological specimens. Patients may also be cross infected with acid fast bacilli (AFB). During a five month period, 12 bronchial wash specimens of 65 patients undergoing bronchoscopy, one bronchoscope and an autocleaner, were contaminated with the AFB, Mycobacterium chelonae (MCH).
Aim: To eradicate AFB contamination of bronchoscopy specimens by identifying sources of contamination and modifying disinfection procedures.
Methods: To identify the source of contamination, samples for AFB culture were taken from three bronchoscopes, the autocleaner and water taps. To eradicate MCH contamination, the bronchoscopes were soaked in 2% glutaraldehyde overnight and flushed with 70% alcohol. Disinfection procedures were altered by using sterile water and containers in cleaning. Autocleaner use was discontinued.
Results: The autocleaner, one bronchoscope and 12 bronchial wash specimens were contaminated with MCH. All contaminants had similar electrophoretic banding on probing of their DNA fragments, suggesting a common clone of origin. After the alterations in disinfection procedures and despite prolonged soaking in 2% glutaraldehyde, three further contaminated wash specimens were isolated from one bronchoscope. Only after ethylene oxide sterilisation of this bronchoscope was the contamination overcome. Since then no further MCH contamination has occurred. No patient required treatment and there has been no clinical evidence of mycobacterial disease.
Conclusion: To avoid contamination of bronchoscopy specimens with MCH, use of autoclearners should be discouraged and sterile water and containers used in cleaning procedures. If MCH contamination occurs in this setting, the bronchoscope and dismantled valve mechanism should undergo ethylene oxide sterilisation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1445-5994.1994.tb01785.x | DOI Listing |
J Prosthet Dent
January 2025
Associate Professor, Department of Dental Medicine, Faculty of Medicine and Dentistry, University of Valencia, Valencia, Spain.
Statement Of Problem: Intraoral scans can be articulated in maximum intercuspal position (MIP) by using an artificial intelligence (AI) based program; however, the impact of edentulous areas on the accuracy of the MIP located using this AI-based program is unknown.
Purpose: The purpose of this in vitro study was to assess the impact of edentulous areas (0, 1, 2, 3, and 4 posterior mandibular teeth) on the accuracy of the MIP located using 3 intraoral scanners (IOSs) and an AI-based program.
Material And Methods: Stone casts articulated in MIP in an articulator were digitized (T710).
Narra J
December 2024
Master of Biomedicine Study Program, School of Medicine and Health Sciences, Universitas Katolik Indonesia Atma Jaya, Jakarta, Indonesia.
Intestinal parasitic infections (IPIs), caused by helminths and/or protozoa, continue to be a significant public health concern in Indonesia. Water access, sanitation, and hygiene practices (WASH) are influential factors for IPIs, especially among children. The aim of this study was to investigate the association between WASH and IPIs among school-aged children.
View Article and Find Full Text PDFBackground And Aims: Intestinal parasitic infection is a common disease that causes significant morbidity and mortality in developing countries, including Ethiopia. The burden of intestinal parasites is worse in schoolchildren, which leads to absenteeism from school, lower cognitive ability, affects academic performance, and causes malnutrition and anemia. While many studies have been reported, there is a paucity of published data in Debre Markos.
View Article and Find Full Text PDFJ Anal Toxicol
January 2025
Arcadia University, Glenside, Pennsylvania, United States.
In postmortem forensic investigation cases where the bladder is voided or dehydrated prior to autopsy, it is possible to wash the bladder with saline and collect the 'bladder wash' and any residual urine for toxicological analysis. While not conventional, this study aims to determine the use of bladder washes as alternative specimens in postmortem forensic toxicology. Comprehensive drug and alcohol analysis was performed on blood, urine, vitreous humor and bladder wash samples.
View Article and Find Full Text PDFJ Prosthet Dent
December 2024
Professor, Department of Restorative Dentistry, University of Washington School of Dentistry, Seattle, Wash. Electronic address:
Statement Of Problem: The angled screw channel (ASC) design has been well accepted for implant prostheses. However, investigation into the behavior of the ASC connection is sparse.
Purpose: The purpose of this in vitro study was to assess the effect of cyclic loading on the internal connection of an ASC system compared with straight access systems by measuring reverse torque values (RTVs) and using microcomputed tomography (µCT) imaging.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!