FDA labeling requirements for disinfection of endoscopes: a counterpoint.

Infect Control Hosp Epidemiol

Department of Medicine, University of North Carolina, Chapel Hill 27599-7030, USA.

Published: April 1995

Endoscopes are used widely for the diagnosis and therapy of medical disorders. To prevent spread of nosocomial infection, all endoscopes should undergo thorough cleaning and high-level disinfection following each use. The Food and Drug Administration (FDA) has approved a user-friendly package label for one liquid chemical germicide that requires a 45-minute immersion at 25 degrees C to support a high-level disinfection label claim. Scientific data reviewed here suggest that one can achieve at least an 8-log reduction in M tuberculosis contamination with cleaning (4 logs) followed by chemical disinfection for 20 minutes (4 to 6 logs). The FDA should modify the label to state that if cleaning is accomplished using a standardized cleaning protocol, then a 20-minute immersion at 20 degrees C will be sufficient to achieve high-level disinfection.

Download full-text PDF

Source
http://dx.doi.org/10.1086/647095DOI Listing

Publication Analysis

Top Keywords

high-level disinfection
12
immersion degrees
8
disinfection
5
fda labeling
4
labeling requirements
4
requirements disinfection
4
disinfection endoscopes
4
endoscopes counterpoint
4
counterpoint endoscopes
4
endoscopes diagnosis
4

Similar Publications

Background: The use of laryngeal masks (LM) has increased widely today, both in anesthesia and in emergency cases. LM are available as reusable and disposable. Although reuse of disposable LM is not recommended, they are reused again after decontamination with different methods in anesthesia units in some countries.

View Article and Find Full Text PDF

Efficacy of cleaning, disinfection and sterilization modalities for addressing infectious drug-resistant fungi: A review.

J Appl Microbiol

January 2025

Centre for Sustainable Disinfection and Sterilization, Technological University of the Shannon, Athlone Campus, N37 HD68, Ireland.

This is a timely and important review that focuses on the appropriateness of established cleaning, disinfection and sterilization methods to safely and effectively address infectious fungal drug-resistant pathogens that can potentially contaminate reusable medical devices used in healthcare environment in order to mitigate the risk of patient infection. The release of the World Health Organisation (WHO) fungal priority pathogen list (FPPL) in 2022 highlighted the public health crisis of antimicrobial resistance (AMR) in clinically relevant fungal species. Contamination of medical devices with drug-resistant fungal pathogens (including those on the FPPL) in healthcare are rare events that are more likely to occur due to cross-transmission arising from lapses in hand-hygiene practices.

View Article and Find Full Text PDF

: Infection prevention and control (IPC) is essential to ensure the safety of dental personnel and patients. This study aimed to assess the knowledge and compliance of dental undergraduate students, interns, and postgraduate students with IPC measures in prosthodontic procedures. : A cross-sectional observational study was conducted at the College of Dentistry, King Saud bin Abdulaziz University for Health Sciences, involving 216 participants selected using stratified random sampling.

View Article and Find Full Text PDF

Introduction/purpose: There are varying international recommendations regarding the minimum level of disinfection required for ultrasound transducers used in percutaneous procedures. While some guidelines recommend high-level disinfection (HLD), others question the additional benefit this delivers over low-level disinfection (LLD).

Methods: This narrative review identifies current guidelines and evaluates the evidence used to support disinfection recommendations for ultrasound transducers used in percutaneous procedures.

View Article and Find Full Text PDF

Introduction: There is increasing prevalence of single-use flexible laryngoscopes in Otolaryngology. This study aims to quantify and compare the environmental outcomes of single-use disposable flexible laryngoscopes (SUD-Ls) and reusable flexible laryngoscope (R-Ls).

Methods: The ISO 14040 standardized Life Cycle Assessment (LCAs) was utilized to estimate the environmental footprint of SUD-L and R-L.

View Article and Find Full Text PDF