AI Article Synopsis

  • A study of 76 PTCA-balloons after coronary angioplasty revealed that 68% had coronary plaque particles on their surfaces, with an average of about 5 particles per balloon.
  • Longer plaque imprints were more common at lower dilatation pressures, while a lower cholesterol and triglyceride level in patients correlated with higher plaque imprints.
  • Energy dispersive x-ray analysis showed significant differences in chemical content between the plaque particles and balloon surfaces, highlighting the potential risks of plaque particles leading to complications like coronary spasms.

Article Abstract

Unlabelled: Seventy-six PTCA-balloons after coronary angioplasty were studied for superficial changes using scanning electron microscopy (SEM) after fixing in glutardialdehyde. Coronary plaque particles were identified on the balloon surface in 52 cases (68%). Twelve new and unused balloons were subjected to the same chemical treatment and SEM showed no imprints. The average length of the longest imprinted plaques was 128 +/- 201 microns and the average number of plaque particles per balloon was 4.9 +/- 2.7. The maximal dilatation pressure and the number of dilatations showed no influence on the impregnation of plaque particles. However, longer plaque imprints tended to occur under low dilatation pressure. Imprints of plaque particles were significantly higher in patients with low cholesterol (p = 0.0001) and low triglycerides (p = 0.0016). No correlation was seen between imprint length and lipid levels. Similarly, the different balloon materials (polyethylene, polyolefincopolymer) showed no significant differences with regard to plaque occurrence. The PTCA-balloons, plaque particles and six coronary plaques obtained after endatherectomy were subjected to energy dispersive x-ray analysis (EDX) under SEM as EDX reveals qualitative and quantitative information about the structural elements. Highly significant differences in calcium, sodium, phosphorus and silicon contents (p = 0.0000) between plaque particles and balloon surface were observed, owing to the absence of these in balloon material. Thus EDX offers additional advantages over SEM in that it clearly differentiates deformed balloon surface, plaque particle, and retained contrast medium.

Conclusion: Plaque particles can be recovered from balloon surfaces after PTCA. Depending upon their size, they could lead to coronary spasm or microembolic phenomenon.

Download full-text PDF

Source

Publication Analysis

Top Keywords

plaque particles
32
balloon surface
16
plaque
11
coronary plaque
8
particles
8
balloon
8
particles balloon
8
dilatation pressure
8
[imprints coronary
4
particles ptca
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!