Polymer coatings have been suggested to decrease the thrombogenicity of metallic intravascular stents. The purpose of the present study was to investigate the intimal response to two different polymers when used as coatings for stents implanted in normal porcine coronary arteries. Non-articulated stainless steel-slotted tube stents were coated with either a biodegradable poly(organo)phosphazene with amino-acid ester side groups or a biostable polyurethane prepared from an amphiphilic polyether, dephenylmethane-4,4'-diisocyanate and butane diol as chain extender. In order to induce vascular wall injury, the stents were deployed using an oversized balloon. At 6 weeks follow-up, the angiographic luminal diameter measured in four polyurethane-coated stents and in six bare metallic stents was similar and 20% less than immediately post-stenting. However, in four polyphosphazene-coated stents the difference was 65% (P = 0.01 when compared to bare metal). At post-mortem morphometry the degree of luminal area stenosis was also similar in polyurethane-coated and in bare metallic stents (32 +/- 7.6% vs. 39 +/- 14%, NS) but reached 81 +/- 19% in polyphosphazene-coated stents (P < 0.03 when compared to bare metal). Thus, poly(organo)phosphazene induced a more pronounced histiolymphocytic and fibromuscular reaction than amphiphilic polyurethane, which appeared to be promising as biocompatible stent coating and, consequently, as a potential carrier for vasoactive drugs.

Download full-text PDF

Source
http://dx.doi.org/10.1016/0021-9150(94)05472-uDOI Listing

Publication Analysis

Top Keywords

metallic stents
12
stents
10
stents implanted
8
implanted normal
8
normal porcine
8
porcine coronary
8
coronary arteries
8
bare metallic
8
polyphosphazene-coated stents
8
compared bare
8

Similar Publications

Concurrent malignant biliary and gastric outlet obstruction requires urgent palliative intervention to improve patient quality of life and permit systemic therapy. Traditional management has been surgical gastrojejunostomy and hepaticojejunostomy, two morbid procedures. Comparatively, endoscopic stenting can relieve both sites of obstruction with less complications and quicker recovery.

View Article and Find Full Text PDF

Suspicion of vascular injury during endoscopic retrograde cholangiopancreatography (ERCP) should be raised in the event of intraprocedural bleeding, persistent hyperbilirubinemia, and sepsis despite biliary stenting. Most inadvertent portal vein (PV) cannulations during ERCP are innocuous, and mere withdrawal of guidewire and catheter suffices. However, unintentional PV stenting, particularly with larger metallic stents, increases the likelihood of significant bleeding.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!