Background: Thirty-eight limbs with iliac occlusive disease were treated with Palmaz stents from 1987 through 1991.

Methods: Indications for stent utilization included dissection induced by percutaneous transluminal balloon angioplasty (PTA) (10), restenosis after PTA (nine), post-PTA residual stenosis (nine), multiple stenoses or occlusion (five), and unfavorable location (five).

Results: The ankle/brachial pressure index increased from 0.53 +/- 0.27 to 0.8 +/- 0.26 after stent deployment. The intraluminal pressure gradient decreased from 31.9 +/- 16.3 to 0.9 +/- 2.2 mm Hg after stent deployment. Complications included pseudoaneurysm (one), arteriovenous fistula (one), iliac perforation (one), groin hematoma (two), and occlusion (two). Follow-up arteriogram showed stenosis proximal or distal (n = 4) or within the stents (n = 4). These were treated with PTA or stents. Two patients required an aortobifemoral graft. Nine patients have died. Life table analysis showed a 1-, 3-, and 5-year primary and secondary cumulative patency of 87% +/- 5.9%, 74% +/- 8.2%, and 63% +/- 10% and 91% +/- 5.1%, 91% +/- 5.6%, and 86% +/- 7.6%, respectively.

Conclusions: Palmaz stents, often required to salvage a PTA failure, appear to maintain overall patency at a high level. However, intimal hyperplasia and the progression of atherosclerotic disease may result in a need for additional procedures to obtain this favorable outcome.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0039-6060(05)80025-xDOI Listing

Publication Analysis

Top Keywords

+/-
10
iliac occlusive
8
occlusive disease
8
palmaz stents
8
stent deployment
8
91% +/-
8
long-term follow-up
4
follow-up palmaz
4
stent
4
palmaz stent
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!