A Monte Carlo computational model of the imaging chain has been used to investigate the performance of x-ray contrast agents with atomic number, Z, 53 < or = Z < or = 90 with respect to physical image quality descriptors (contrast and signal to noise ratio, SNR) and patient mean absorbed dose. Contrast agents of equal molar concentrations were used within a water slab (simulating the patient). The imaging conditions were chosen to represent adult and paediatric examinations. For all tube potentials studied (40-140 kV), the contrast agents with the highest atomic numbers (bismuth and thorium) gave the highest contrast. In analogue screen-film imaging, several other contrast agents could produce a higher image contrast than iodine in a limited range of tube potentials. This advantage could alternatively be effected as a reduced amount of administered contrast agent, or as a reduced mean absorbed dose in the patient. In digital imaging, a lower mean absorbed dose for a constant SNR than that with iodine can be achieved for ranges of tube potentials and contrast agents. Bismuth and thorium yield a lower dose than iodine at all studied tube potentials. Gadolinium and erbium could alternatively be used at a broad range of tube potentials above 90 kV with a dose penalty of only 5-20%.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/40/7/005DOI Listing

Publication Analysis

Top Keywords

contrast agents
24
tube potentials
20
absorbed dose
12
contrast
10
performance x-ray
8
x-ray contrast
8
monte carlo
8
model imaging
8
imaging chain
8
bismuth thorium
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!