Performance of a 3.2-mm to 6-mm adaptor.

Pacing Clin Electrophysiol

Robert L. Batey Cardiology Center, Pacemaker Systems Technology School, Bradenton, Florida 34209.

Published: February 1994

A study was undertaken to evaluate the performance of a 366-08 adaptor that adapts an in-line bipolar 3.2-mm connector to a 6-mm pulse generator connector. A total of 14 adaptors were implanted in individual patients. Eight of the 14 adaptors failed, an additional failure is probable. The average follow-up time was 32.1 months. The most likely cause of failure was the spring-loaded mechanism of the adaptor, which fatigues with time. All eight of the documented failures presented with loss of capture and were accompanied by increased measured lead impedances. In addition, some failed systems demonstrated loss of sensing. As a result, prophylactic replacement of ventricular lead systems incorporating this adaptor would be advisable. Otherwise, more frequent monitoring with lead impedance measurements may suffice.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.1994.tb01363.xDOI Listing

Publication Analysis

Top Keywords

performance 32-mm
4
32-mm 6-mm
4
adaptor
4
6-mm adaptor
4
adaptor study
4
study undertaken
4
undertaken evaluate
4
evaluate performance
4
performance 366-08
4
366-08 adaptor
4

Similar Publications

Purpose: High dose rate (HDR) prostate brachytherapy (BT) procedure requires image-guided needle insertion. Given that general anesthesia is often employed during the procedure, minimizing overall planning time is crucial. In this study, we explore the clinical feasibility and time-saving potential of artificial intelligence (AI)-driven auto-reconstruction of transperineal needles in the context of US-guided prostate BT planning.

View Article and Find Full Text PDF

Severe mitral regurgitation (MR) following surgical repair of the mitral valve poses a significant clinical challenge. Patients who have undergone surgery are typically at high risk for a second operation. This report details the case of a 54-year-old male who underwent aortic valve replacement and mitral valve repair using a 34-ring, 14 years prior.

View Article and Find Full Text PDF

: As Repeated Low-Level Red Light (RLRL) therapy is becoming increasingly prevalent in clinical practice, mainly in the Far East, largely due to its child-friendly nature and the feasibility of home use, this study aims to conduct a systematic review and meta-analysis to evaluate the efficacy of RLRL therapy in managing childhood myopia, specifically in relation to axial length (AL) and spherical equivalent refraction (SER), across a larger group of children aged from 6 to 16 years. : A systematic literature search was performed using PubMed, Scopus, and Web of Science to access relevant databases and to locate outcome studies. Eligibility criteria included publication type, participant characteristics, and outcomes report.

View Article and Find Full Text PDF

Background And Objectives: Brain tumors are complex diseases that require careful diagnosis and treatment. A minor error in the diagnosis may easily lead to significant consequences. Thus, one must place a premium on accurately identifying brain tumors.

View Article and Find Full Text PDF

Transjugular transcatheter edge-to-edge mitral valve repair in a patient with functional mitral regurgitation: a case report.

Eur Heart J Case Rep

January 2025

Department of Cardiovascular Medicine, Sendai Kousei Hospital, 1-20 Tsutsumidori-amamiya, Aoba Ward, Sendai, Miyagi 9810914, Japan.

Article Synopsis
  • The transcatheter edge-to-edge mitral valve repair (M-TEER) using the MitraClip system is usually done via the transfemoral approach, but the transjugular approach can be a viable alternative when the former isn't possible.
  • A 57-year-old man with heart failure and severe functional mitral regurgitation could not undergo the transfemoral approach due to an occluded inferior vena cava, leading to the decision to use the transjugular technique instead.
  • The procedure was successful with no complications, significantly reducing the patient's mitral regurgitation, and he showed sustained improvement a year later.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!