A heterologous RIA system was set up for measuring bovine, ovine and porcine calcitonin (CT). The system consisted of porcine GT used as standard and for the preparation of an iodinated tracer. The antiserum used was raised against ovine GT. For each analysis was used 25–200 µl blood plasma. Practical detection limit was 0.25 µg of CT per litre of blood plasma. The parallelism between the dose response curves for the p-GT standard and for the assay of increasing amounts of bovine, ovine and porcine blood plasma showed the suitability of the present assay system to study the GT secretion in these species. Furthermore, the reliability of the method was verified by a clearly recognized CT response to calcium infusion.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8317743PMC
http://dx.doi.org/10.1186/BF03546880DOI Listing

Publication Analysis

Top Keywords

bovine ovine
12
ovine porcine
12
blood plasma
12
evaluation radioimmunoassay
4
radioimmunoassay technique
4
technique measuring
4
measuring calcitonin
4
calcitonin bovine
4
ovine
4
porcine
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!