An acrylic head phantom was irradiated during a computed tomographic scan with four commercial scanners. Measurements of the spatial distribution of the radiation dose on the surface and internal to the phantom were performed for the scan plane and the scattered beam at various distances from the scan plane. The surface scatter dose was found to be considerably smaller than that for internal scatter. A significant increase in radiation exposure within the head phantom due to internal radiation scatter, and an asymmetrical primary beam profile for dual slice scanners were also noted.