Unlabelled: The frictional torque of the Trapezoidal-28 total hip replacement and the Tharies SR-3 and SR-5 surface replacements were measured in the University of California at Los Angeles hip-joint simulator in calf serum. Loads of as much as 890 newtons were applied while the femoral component oscillated through a 60-degree arc at forty cycles per minute. Frictional torque, as measured by a ring transducer at a fixed distance from the center of rotation, was seen to diminish as the thickness of the polyethylene increased and was found to be relatively proportional to the diameter of the femoral component. The torques measured in this study were found to be twenty to 100 times smaller than the static torques to failure for acetabular fixation that have been reported in the literature. Finally, no permanent deformation of the thin acetabular shells attributable to the cementing process was apparent when the recommended technique of surgical insertion was utilized.

Clinical Relevance: Surface replacement components have been shown to generate greater frictional torques at the acetabular bone-cement interface than do smaller-diameter conventional hip replacements. This indicates that the local mechanical-loading conditions of the hip joint are not the same for both types of replacements, and this should be considered when selecting the type of arthroplasty for an individual patient.

Download full-text PDF

Source

Publication Analysis

Top Keywords

frictional torque
12
conventional hip
8
hip replacement
8
femoral component
8
frictional
4
torque surface
4
surface conventional
4
hip
4
replacement unlabelled
4
unlabelled frictional
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!