A method for computed tomography using equipment available on some radiotherapy simulators has been investigated. An offset narrow fan beam of X rays passes transversely through the object and is continuously detected by an image intensifier and TV camera system while the beam and detector complete a rotation about the object in 30 seconds. Several TV lines of the images obtained at equal angular increments are digitised and summed to produce transmission projections from which a transverse section is formed by filtered back-projection. The dynamic range required of the camera is reduced by modifying the beam with a shaped metal filter. Images are produced without logarithmic processing or X-ray output calibration. Digital control circuitry is used for digitisation and spatial filtering, and a high-speed arithmetic array processor and digital frame store are used for back projection and image storage (Cynosure Imaging Systems Ltd.). Phantom studies using an object on a turntable with a stationary X-ray tube and intensifier showed the system to be capable of good spatial resolution and reasonable density resolution. Acceptable images of humans have been produced with the equipment added to a Ximascan simulator (TEM Instruments Ltd.). Radiation dose to the irradiated section can be as little as 200 microGy (20 mrad).

Download full-text PDF

Source
http://dx.doi.org/10.1259/0007-1285-57-673-47DOI Listing

Publication Analysis

Top Keywords

image intensifier
8
intensifier camera
8
investigations x-ray
4
x-ray image
4
camera imaging
4
imaging transverse
4
transverse sections
4
sections humans
4
humans method
4
method computed
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!