The authors have tested a new device that records the filtration pressure of a suspension of red blood cells (R.B.C.) passing through a 3 micron pore diameter filter. As the flow rate is constant, the pressure increase allows to appreciate an index of the R.B.C. deformability. In order to test the performances of this new equipment, the authors have studied the influence of the variation of the following parameters. Haematocrit from 0.5 to 2%, Buffer composition; time elapsing from drawing to measurement: from hour 0 to hour 3. This analysis has been performed on donors; on packed red cells preserved in CPD, from D0 to D21 and on patients specimens and the results have been compared to the whole blood filtration according to the Reid and Dormandy method. The following parameters were simultaniously measured: Whole blood viscosity by means of a Low Shear 30 viscosimeter, Intraerythrocyte ATP by bioluminescence, Membrane lipid composition (cholesterol, phospholipids), Scanning electron microscopy. This new apparatus, measuring the filtration pressure seems to be a more sensitive and reproductible method allowing to approach the RBC deformability, and quite usefull in clinical haemorheology.

Download full-text PDF

Source
http://dx.doi.org/10.3233/bir-1984-23s145DOI Listing

Publication Analysis

Top Keywords

filtration pressure
12
red blood
8
rbc deformability
8
filtration
4
pressure red
4
blood
4
blood cell
4
cell deformability
4
deformability evaluation
4
evaluation device
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!