An improved simultaneous tomography cassette according to P. Landau was tried out for four months using four tomographs in routine work. The mode of operation is based on accurate control of the relative speeds of the individual x-ray films resulting in simultaneous imaging of 6 equidistant tomographic levels. Clinical testing was effected in 80 cases: nephrotomography, of the lungs, the hilum, and the skeleton. In particular, the article describes imaging of the renal arteries by simultaneous tomography for the purpose of finding out the cause of hypertension, and if there is suspicion of a space-occupying growth in the kidney, basing on the urogram. The specific advantages of this technique are, on the one hand, improved diagnostic efficiency (the tomograms are taken during the same respiratory phase, more rapid diagnosis especially with accident patients), and, on the other hand, an important reduction in the x-ray exposure of the patient; furthermore, the life of the x-ray tube is prolonged, and there is a definite saving of time for both patient and personnel, the image quality being comparable with that of single-layer tomography.