Measurements of gas exchange have been demonstrated to be clinically useful in the care of critically ill and malnourished patients. Using principles of indirect calorimetry, resting energy expenditure (REE) can be calculated from gas exchange data and used as the basis for designing a nutritional support regimen as well as for following the patient's metabolic state. This study demonstrates that a relatively minor procedure, such as percutaneous muscle biopsy, can induce temporary but major increases in gas exchange and lead to an overestimation of REE. Four studies were performed on 3 healthy adult subjects admitted to the Surgical Metabolism Unit for nutritional study. A percutaneous muscle biopsy was performed with the subject inside a canopy with continuous recording of oxygen consumption (VO2) and carbon dioxide production (VCO2). After the muscle biopsy, VCO2 and VO2 increased 93 and 103% (at their peak value), respectively. The mean duration that these changes persisted at least 15% above control was 10.6 +/- 7.8 (SD) and 11.4 +/- 5.9 min of VCO2 and VO2, respectively. Thus, considerable artifacts in the estimation of REE can occur due to painful stimuli.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00003246-198309000-00016DOI Listing

Publication Analysis

Top Keywords

gas exchange
12
muscle biopsy
12
resting energy
8
energy expenditure
8
percutaneous muscle
8
vco2 vo2
8
artifacts measurement
4
measurement resting
4
expenditure measurements
4
measurements gas
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!