Presently, patient exposure and x-ray tube kilovoltage are determined by image visibility requirements on x-ray film. With the employment of image-processing techniques, image visibility may be manipulated and the exposure may be determined only by the desired information content, i.e., by the required degree of tissue-density descrimination and spatial resolution. This work gives quantitative relationships between the image information content and the patient exposure, give estimates of the minimum exposures required for the detection of image signals associated with particular radiological exams. Also, for subject thickness larger than approximately 5 cm, the results show that the maximum information content may be obtained at a single kilovoltage and filtration with the simultaneous employment of image-enhancement and antiscatter techniques. This optimization may be used either to reduce the patient exposure or to increase the retrieved information.