We have evaluated a recently introduced colour test kit for the determination of serum angiotensin I-converting enzyme catalytic activity. p-Hydroxyhippuric acid, liberated from p-hydroxyhippuryl-L-histidyl-L-leucine by angiotensin I-converting enzyme, is converted into a quinoneimine dye with an absorption maximum at 505 nm. The procedure shows excellent linearity over the whole range of catalytic activities found in serum. Intra- and inter-assay coefficients of variation are 2-5 and 7-10% respectively. Correlation with a modified Cushman-Cheung ((1971) Biochem. Pharmacol. 20, 1637-1648) method currently used in our laboratory is good, with r = 0.985 and a regression equation of y (colour kit) = 0.423 x (Cushman-Cheung) + 0.765. Haemoglobin, lipids, bilirubin and prednisone do not interfere but uric acid in concentrations higher than 600 mumol/l does. No extraction step is required. The assay is very rapid, and more than twenty samples can be determined in an hour.

Download full-text PDF

Source
http://dx.doi.org/10.1515/cclm.1983.21.12.845DOI Listing

Publication Analysis

Top Keywords

angiotensin i-converting
12
i-converting enzyme
12
test kit
8
evaluation test
4
kit rapid
4
rapid simple
4
simple colorimetric
4
colorimetric measurement
4
measurement angiotensin
4
enzyme serum
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!