The roentgenographic identification of femoral component loosening after hip surface arthroplasty is often impossible because the metallic femoral component obscures the bone-cement interface. The use of combined technetium sulfur colloid and technetium methylene diphosphonate radionuclide imaging has been especially useful in the diagnosis of loosening. In 40 patients, follow-up combined TcSC and TcmDP scans at an average of three, nine, and 27 months postoperation revealed significant differences in the isotope uptakes in patients who had loose prostheses compared with those without complications. Scans were evaluated by first dividing them into eight anatomical regions and then rating the uptake in each region or 'zone' on a five-point scale. Results were compared using the Student's t-test and differences were noted between normal controls and patients who had femoral component loosening. Combining both TcSC and TcmDP studies increased the statistical significance obtained when comparing patients who had complications to those in the control group.

Download full-text PDF

Source

Publication Analysis

Top Keywords

femoral component
12
hip surface
8
surface arthroplasty
8
radionuclide imaging
8
component loosening
8
tcsc tcmdp
8
identification hip
4
arthroplasty failures
4
failures tcsc/tcmdp
4
tcsc/tcmdp radionuclide
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?