At the present time, the relatively high atrial stimulation threshold and the low and often unstable P wave amplitude are the main concerns in atrial pacing. As a further contribution to the solution of this problem we used a screw-in atrial lead in 16 patients. During the implantation, atrial mapping was performed in order to establish the position with the best electrical performance. The acute threshold for voltage at a pulse duration of 1 ms averaged 0.49 +/- 0.18 Volt. The acute peak-to-peak P wave amplitude averaged 4.6 mV +/- 1.3. No complication occurred at the time of the implantation. Follow-up periods averaged 12 months. No dislocation or bleeding was observed during a follow-up period of up to 28 months. Using this lead together with intraoperative atrial mapping, it is possible to achieve an acute strength-duration relationship in the atrium at a level similar to that in the ventricle. The results of this study demonstrate that intraoperative mapping is an important stein in enhancing the safety margins and long-term performance of atrial pacing.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-2007-1022446DOI Listing

Publication Analysis

Top Keywords

atrial pacing
12
wave amplitude
8
atrial mapping
8
atrial
7
improved technique
4
technique permanent
4
permanent atrial
4
pacing time
4
time high
4
high atrial
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!