Background: The use of capturing devices may become required for the continued use desflurane. We tested the percentage of desflurane captured by a charcoal filter (CONTRAfluran)-workstation (Aisys) combination in vitro.
Methods: Desflurane in O2/air was administered via an Aisys workstation into a 2 L test lung that was insufflated with CO2 (160 mL/min). First, to confirm all vaporized desflurane reached the capturing device, the amount of desflurane collected in a Douglas bag attached to the machine exhaust was compared to the vaporized amount during 15 min runs with the following fresh gas flow (FGF)//vaporizer setting combinations: (L/min // %): 0.3//8, 0.5//8, 1//6, 2//6, 3//6, 4//6, 5//6, and 6//6. Next, to determine the effect of CO2, the capturing device weight gain was measured with the same FGF ran over 1 hour but without desflurane. Finally, the ratio of the capturing device weight gain / vaporizer weight loss (= performance, expressed in %) was determined for the same 15 min runs with desflurane vaporizer settings described above. All experiments were arbitrarily repeated five times.
Results: The amount of vaporized desflurane did not differ from the amount collected in the Douglas bag. When CO2, O2, and air were delivered without desflurane, the capturing device lost a relatively small amount of weight (<5 g), especially with FGF ≤ 1 L/min. Finally, performance with 0.3, 0.5-2 and 3-6L/min FGF was 103, 100, and 95 - 93 %, respectively.
Conclusions: CONTRAfluran charcoal filter in vitro performance for desflurane in O2/air combined with the Aisys workstation ranged from 93-103 % with FGF of 0.3 to 3-6 L/min with vaporizer settings that reflect clinical conditions. Defining the place of charcoal filters in clinical practice requires full life cycle analysis of both the charcoal and inhaled agent.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/ALN.0000000000005445 | DOI Listing |
Anesthesiology
March 2025
Professor of Anesthesiology and Perioperative Medicine, Department of Basic and Applied Medical Sciences, Ghent University, Ghent, Belgium; Professor, Department of Anesthesiology, UZLeuven, Leuven, Belgium & Department of Cardiovascular Sciences, KULeuven, Leuven, Belgium; Staff anesthesiologist, Department of Anesthesiology, OLV Hospital, Aalst, Belgium.
Background: The use of capturing devices may become required for the continued use desflurane. We tested the percentage of desflurane captured by a charcoal filter (CONTRAfluran)-workstation (Aisys) combination in vitro.
Methods: Desflurane in O2/air was administered via an Aisys workstation into a 2 L test lung that was insufflated with CO2 (160 mL/min).
ACS Nano
March 2025
State Key Lab for Mesoscopic Physics and Frontiers Science Center for Nano-optoelectronics, Electron Microscopy Laboratory, School of Physics, Peking University, Beijing 100871, China.
The solid-state nanopore technique holds the potential to develop mechanically stable and miniaturized DNA sequencing devices. However, the limited temporal resolution due to the high electric field inside the nanopore and the lack of an effective speed control strategy have hindered the realization of sequencing. Here, we reported a quad-array (four nanopores milled with ∼30 nm interpore spacing as a detection unit) that induced a redistribution of the electric field inside and outside the nanopore array and offered high-resolution discrimination of four ssDNA homopolymer types.
View Article and Find Full Text PDFAdv Mater
March 2025
Walker Department of Mechanical Engineering, The University of Texas at Austin, Austin, TX, 78712, USA.
Lithium-ion batteries are indispensable power sources for a wide range of modern electronic devices. However, battery lifespan remains a critical limitation, directly affecting the sustainability and user experience. Conventional battery failure analysis in controlled lab settings may not capture the complex interactions and environmental factors encountered in real-world, in-device operating conditions.
View Article and Find Full Text PDFCatheter Cardiovasc Interv
March 2025
Division of arrhythmology, San Raffaele Hospital, Milan, Italy.
Background: Performing a left atrial appendage occlusion (LAAO) or catheter ablation with left-sided intracardiac thrombus is considered very-high risk for periinterventional stroke. Cerebral embolic protection (CEP) devices are designed to prevent cardioembolic stroke and have been widely studied in TAVR procedures. However, their role in LAAO and catheter ablation of ventricular tachycardia (VT) or in pulmonary vein isolation (PVI) with cardiac thrombus present remains unknown.
View Article and Find Full Text PDFJ Biomed Opt
March 2025
University of Minnesota, Department of Biomedical Engineering, Minneapolis, Minnesota, United States.
Significance: We introduce a visible-light polarization-sensitive optical coherence tomography (PS-OCT) system that operates in the spectral domain with balanced detection (BD) capability. While the BD improves the signal-to-noise ratio (SNR), the use of shorter wavelengths improves spatial resolution and birefringence sensitivity.
Aim: We aim to implement a new optical design, characterize its performance, and investigate the imaging potential for biological tissues.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!