Background: EDTA-dependent pseudothrombocytopenia (EDTA-PTCP) is an phenomenon that may lead to expensive, time-consuming, and invasive diagnostic procedures as well as unnecessary patient treatment. The purpose of this study was to explore the effects of time, anticoagulant and detection channel on the platelet (PLT) count of EDTA-PTCP samples, and to suggest a better method for correcting spurious low PLT counts.

Methods: In this study, 43 identified EDTA-PTCP samples were collected. The Sysmex XN-9100, Mindray BC-6900 and Mindray BC-5390 haematology analysers were used to test these EDTA-PTCP samples on the following detection channels at different time points: PLT count by impedance method (PLT-I), PLT count by optical method (PLT-O) and PLT count by fluorescent staining (PLT-F).

Results: EDTA-PTCP was time-dependent and small PLT agglutination occurred in most of the corresponding citrate-treated samples. Our results further demonstrated that the detection channel significantly affected the PLT count of the EDTA-PTCP samples. The XN-9100 PLT-F channel exhibited a greater dissociative effect than the XN-9100 PLT-I and PLT-O channels. Moreover, blood samples processed in the PLT-O channel of the Mindray hematology analyzer showed the highest PLT count in EDTA-K2 tubes compared to the other detection channels.

Conclusion: Our data showed that time, anticoagulant and detection channel significantly affected the PLT count in the EDTA-PTCP samples. For the EDTA-PTCP samples, the simplest retest method was to use the PLT-O channel of the Mindray automatic blood analyser within 30 min. In addition, changing the sodium citrate anticoagulant and using the XN-9100 PLT-F channel within 15 min were also suitable for correcting the spurious low PLT of the EDTA-PTCP samples.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11887562PMC
http://dx.doi.org/10.7717/peerj.19103DOI Listing

Publication Analysis

Top Keywords

plt count
28
edta-ptcp samples
28
detection channel
16
time anticoagulant
12
anticoagulant detection
12
count edta-ptcp
12
plt
10
edta-ptcp
9
samples
9
effects time
8

Similar Publications

The present study is the first from Pakistan being reported with an objective to assess performance of Rayto RT-7600Vet hematology analyzer (HA) for Cholistani cattle blood (n = 134), in comparison to the manual hematological methods. The four hematological attributes viz. total erythrocyte count (TEC), hemoglobin (Hb), packed cell volume (PCV) and platelet count (PLT) were deduced through HA (A) and manual (M) methods.

View Article and Find Full Text PDF

Background: This study investigated the expression and clinical significance of coiled-coil domain containing 12 (CCDC12) in the initial diagnosis of acute myeloid leukemia (AML).

Methods: A total of 80 AML patients were enrolled as the experimental group, and 20 normal bone marrow specimens were used as the control group. Clinical data of AML patients were collected.

View Article and Find Full Text PDF

Neutrophil Extracellular Traps Enhance Procoagulant Activity and Predict Poor Prognosis in Patients With Metastatic Breast Cancer.

Int J Gen Med

March 2025

Key Laboratory of Breast Cancer Diagnosis and Treatment Research of Guangxi Department of Education, Nanning, Guangxi, People's Republic of China.

Objective: Neutrophil extracellular traps (NETs) are associated with poor prognosis and an increased risk of venous thromboembolism (VTE) in metastatic breast cancer (MBC). This study aims to determine whether NETs promote hypercoagulability and if NETs and plasma hypercoagulability markers are biomarkers of survival in MBC.

Methods: Circulating levels of neutrophil extracellular trap (NET) markers and hypercoagulability markers (TAT, fibrinogen, and D-dimer) were assessed in 112 MBC patients before treatment, compared to 55 healthy controls.

View Article and Find Full Text PDF

Background: EDTA-dependent pseudothrombocytopenia (EDTA-PTCP) is an phenomenon that may lead to expensive, time-consuming, and invasive diagnostic procedures as well as unnecessary patient treatment. The purpose of this study was to explore the effects of time, anticoagulant and detection channel on the platelet (PLT) count of EDTA-PTCP samples, and to suggest a better method for correcting spurious low PLT counts.

Methods: In this study, 43 identified EDTA-PTCP samples were collected.

View Article and Find Full Text PDF

Introduction: To investigate the effect of continuous renal replacement therapy (CRRT) in patients with severe acute kidney injury (AKI) by using Nafamostat Mesilate (NM).  Methods. Eighty patients with AKI who underwent CRRT from March 2022 to January 2022 were divided into control group (n = 40, treated with unfractionated heparin) and Observation group (n = 40, treated with NM).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!