Objectives: There is a need to better inform clinicians and decision-makers in primary or community care settings on selecting the appropriate point-of-care tests (POCTs) for screening purposes (as a part of the NHS Health Check Programme). Here we provide an overview of the published analytic validity and diagnostic accuracy studies on POCTs for measuring blood lipids that are available on the UK market to determine whether they meet the accuracy specifications based on the 1995 US National Cholesterol Education Program (NCEP) recommendations.
Design: Rapid review of analytical validity and diagnostic accuracy studies.
Data Sources: On 12 May 2023, Medline and Embase were searched. Google Scholar was manually scrutinised to identify additional studies. Key article reference lists were also hand-searched.
Eligibility Criteria: We included analytical validity and diagnostic accuracy studies that compared POCT to laboratory testing (or another POCT) performance for measuring at least total cholesterol (TC) and high-density lipoprotein cholesterol (HDL-C).
Data Extraction And Synthesis: Identified studies were independently reviewed by two researchers using standardised methods of screening. Where necessary, conflicts were resolved by a third reviewer. Title and abstract as well as full texts were screened using prespecified inclusion and exclusion criteria. The quality of identified studies was assessed using QUADAS-2 for diagnostic accuracy studies and a modified quality appraisal tool for studies of diagnostic reliability (QAREL) for analytical validity studies. We assessed the quality of analytical and diagnostic accuracy studies and compared the accuracy of the POCTs for TC, triglyceride (TG), HDL-C and low-density lipoprotein cholesterol (LDL-C) against NCEP standards for mean per cent bias, coefficient of variation or total error. We narratively synthesised analytical and clinical validity evidence from retrieved studies.
Results: This study examined analytical and diagnostic accuracy evidence for the selected POCTs. Through the review of 22 studies, 6 POCTs were identified. All retrieved studies were analytical validity assessments, while five of them also reported diagnostic accuracy information. The majority of evidence focused on Cholestech LDX, CardioChek PA and Accutrend Plus. Evidence of between and within-study heterogeneity was found. Precision measures often showed systematic differences between the POCT and reference standards. Most devices, except for Elemark, met at least one NCEP standard for either TC, TG, HDL-C, or LDL-C.
Conclusions: We found that evidence for two of the devices mostly met the requirements of the NCEP standard of evidence for bias and precision and could be recommended to general practitioners to use in the NHS Health Check programme. These were the Cholestech LDX and the Cobas b101 system.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11883607 | PMC |
http://dx.doi.org/10.1136/bmjopen-2023-080726 | DOI Listing |
Anal Chem
March 2025
College of Chemistry, Jilin Province Research Center for Engineering and Technology of Spectral Analytical Instruments, Jilin University, Qianjin Street 2699, Changchun 130012, China.
The lack of precise, real-time analytical tools for monitoring tumor microenvironment changes during treatment hinders advancements in integrated diagnostic and therapeutic platforms. Traditional caspase-3 monitoring strategies are limited by their inability to address drug resistance and newly discovered apoptotic pathways, leading to reduced accuracy and practicality. To overcome these limitations, we developed a fluorescence-based "Trojan horse" nanosystem, PFpR@CM, featuring high-sensitivity Caspase-1 detection, tumor-targeted delivery, and photothermal therapy.
View Article and Find Full Text PDFTrans R Soc Trop Med Hyg
March 2025
Molecular Epidemiology department, ICMR-National Institute of Malaria Research, Sector 8, Dwarka, 110077 New Delhi, India.
Background: Rapid diagnostic tests (RDTs) are vital for malaria diagnosis, especially in resource-limited areas. RDTs targeting histidine-rich protein 2 (PfHRP2) and its structural homologue PfHRP3 are commonly used for detecting Plasmodium falciparum. However, genetic deletions in these proteins can affect test accuracy.
View Article and Find Full Text PDFChatGPT and other artificial intelligence (AI) tools can modify nutritional management in clinical settings. These technologies, based on machine learning and deep learning, enable the identification of risks, the proposal of personalized interventions, and the monitoring of patient progress using data extracted from clinical records. ChatGPT excels in areas such as nutritional assessment by calculating caloric needs and suggesting nutrient-rich foods, and in diagnosis, by identifying nutritional issues with technical terminology.
View Article and Find Full Text PDFBackground: The Sysmex DI-60 system (Sysmex, Kobe, Japan) is an automated cell image analyzer that captures and analyzes cell images. The aim of this study was to evaluate the performance of DI-60 for leukocyte differential count in comparison with manual differential count and XN-20.
Methods: A total of 205 samples were analyzed and the agreement between DI-60 pre-classification and post-verification by medical technicians was determined.
Background: Tuberculosis is still a significant global health challenge, with latent tuberculosis infection (LTBI) posing a risk for the development of active disease. This study aimed to compare the performance of the STAN-DARD F TB-Feron FIA test with the QuantiFERON-TB Gold Plus, a test endorsed by the World Health Organization (WHO), in detecting LTBI.
Methods: We included 137 participants, out of which 58.