Background: Clear representation of anatomy is essential in the assessment of pathology in computed tomography (CT). With the introduction of photon-counting CT (PCCT) and more advanced iterative reconstruction (IR) algorithms into clinical practice, there is potential to improve low-contrast detectability in CT protocols. As such, it is necessary to perform task-based assessments to optimize protocols and compare image quality between PCCT and energy-integrating CT, like dual-energy CT (DECT) and single-energy CT (SECT).

Purpose: This work aimed to assess low-contrast detectability in abdominal protocols used in clinical PCCT, DECT, and SECT, using both model and human observers.

Methods: Data were acquired with the standard resolution scan mode on a PCCT (NAEOTOM Alpha, Siemens Healthineers, Forchheim, Germany) and a DECT/SECT (SOMATOM Force, Siemens Healthineers, Forchheim, Germany). Detectability was investigated in the CTP 515 low-contrast module of the Catphan 600 phantom, which was surrounded by a fat annulus to simulate an abdomen and resulted in a water equivalent diameter of 298 mm. Supra-slice contrast rods with a nominal 1.0% contrast and diameters of 4, 6, 9, and 15 mm were used. Factory abdominal protocols were adjusted to acquire images with various tube potentials (70, 90, 120, and 140 kV in PCCT; 70/150Sn and 80/150Sn kV in DECT; 100 and 120 kV in SECT), virtual monoenergetic image (VMI) energy levels (40 to 140 keV in PCCT and DECT), doses (5, 10 mGy in PCCT; 10 mGy in DECT and SECT), and IR settings (Br40 kernel, no quantum IR (QIR) and QIR levels 1 to 4 in PCCT; advanced modeled IR (ADMIRE) level 3 in DECT and SECT). Mixed DECT (linear blending of the images at two tube voltages) images were also reconstructed. The noise power spectrum and task transfer function of each scan protocol were quantified; the detectability index for each protocol was also determined using in-house implementations of model observers (non-prewhitening matched filters with internal noise, NPWI, and with an eye filter and internal noise, NPWEI) and human observers (in-house four-alternative forced choice, scoring with 95% confidence intervals).

Results: Results show that the image noise is minimized at a VMI energy corresponding to the applied spectrum's mean energy in PCCT and with VMI settings of 70 and 80 keV for 70/150Sn and 80/150Sn tube potential pairs, respectively, in DECT. With respect to the human observer detectability index calculations, the normalized root-mean-square error for the NPWI and NPWEI model observers was 5% and 12%, respectively. PCCT VMI improves low-contrast detectability. Additionally, detectability can be matched between PCCT protocols by increasing the QIR strength level when reducing the dose. Not only does PCCT VMI outperform DECT VMI, but also DECT VMI outperforms DECT mixed imaging in improving low-contrast detectability.

Conclusions: Low-contrast detectability is optimized when the appropriate VMI energy level is selected in PCCT and DECT to minimize image noise. PCCT improves low-contrast detectability and may allow for dose reduction in abdominal protocols compared to both DECT and SECT. The non-prewhitening model observer with internal noise better quantified low-contrast detectability without the inclusion of an eye filter.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.17717DOI Listing

Publication Analysis

Top Keywords

low-contrast detectability
28
dect sect
20
pcct dect
16
pcct
15
dect
14
abdominal protocols
12
vmi energy
12
internal noise
12
pcct vmi
12
detectability
11

Similar Publications

Mixture of Expert-Based SoftMax-Weighted Box Fusion for Robust Lesion Detection in Ultrasound Imaging.

Diagnostics (Basel)

February 2025

Department of Electrical and Computer Engineering, College of Information and Communication Engineering, Sungkyunkwan University, Suwon 440-746, Republic of Korea.

Ultrasound (US) imaging plays a crucial role in the early detection and treatment of hepatocellular carcinoma (HCC). However, challenges such as speckle noise, low contrast, and diverse lesion morphology hinder its diagnostic accuracy. To address these issues, we propose CSM-FusionNet, a novel framework that integrates clustering, SoftMax-weighted Box Fusion (SM-WBF), and padding.

View Article and Find Full Text PDF

Rotator cuff tendon tears, the most common shoulder injuries, are typically diagnosed mainly through MRI, but can also be seen on ultrasound (US), a much less costly test that currently requires highly-trained human expert operators. An AI tool to identify full-thickness rotator cuff tears in US could make this test much more accessible in clinical practice. We propose a two-step approach starting with segmentation and followed by classification.

View Article and Find Full Text PDF

High-attenuation (HA) artifacts may lead to obscured subtle lesions and lesion over-estimation in digital breast tomosynthesis (DBT) imaging. High-attenuation artifact suppression (HAAS) is vital for widespread DBT applications in clinic. The conventional HAAS methods usually rely on the segmentation accuracy of HA objects and manual weighting schemes, without considering the geometry information in DBT reconstruction.

View Article and Find Full Text PDF

Background: Clear representation of anatomy is essential in the assessment of pathology in computed tomography (CT). With the introduction of photon-counting CT (PCCT) and more advanced iterative reconstruction (IR) algorithms into clinical practice, there is potential to improve low-contrast detectability in CT protocols. As such, it is necessary to perform task-based assessments to optimize protocols and compare image quality between PCCT and energy-integrating CT, like dual-energy CT (DECT) and single-energy CT (SECT).

View Article and Find Full Text PDF

Deep Learning-Driven Automated High-Content dSTORM Imaging with a scalable Open Source Toolkit.

Biophys Rep (N Y)

February 2025

Rudolf Virchow Center for Integrative and Translational Bioimaging, University of Würzburg, Würzburg, Germany. Electronic address:

Super-resolution microscopy offers the ability to visualize molecular structures in biological samples with unprecedented detail. However, the full potential of these techniques is often hindered by a lack of automated, user-independent workflows. Here, we present an open-source toolkit that automates dSTORM super-resolution microscopy using deep learning for segmentation and object detection.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!