Purpose: Japanese normal databases (NDB) for quantitative evaluation of Tc myocardial perfusion SPECT have traditionally been created by the filtered back projection (FBP) method but are also applied to the iterative reconstruction (IR) method in some institutions. Although creating the NDB specifically for the IR method is desirable, it has not been done partly due to the ethical aspects of creating an institution-specific database. This study investigated the impact of using NDBs created by the FBP method on quantitative evaluations performed with the IR method.

Methods: We analyzed single-photon emission computed tomography (SPECT) images from patients undergoing myocardial perfusion study in two hospitals with three SPECT vendors. Images were processed with and without various corrections, and a polar map was created by QPS software using a 17-segment model to obtain %uptake and scores. The effects of the reconstruction and correction methods were evaluated to determine the difference in the number of local segmental analyses by coronary artery territories.

Results: In most segments, quantitative values showed no significant difference across device types, with or without corrections. However, both choices of the SPECT device and correction methods influenced the scores in specific coronary artery regions. Application of attenuation correction resulted in lower defect scores in the right coronary artery region.

Conclusion: Applying FBP-derived NDB to the IR method generally yielded a consistent patient diagnosis in terms of defect score analysis. However, since the quantitative evaluation was partly affected by various correction methods and devices, recognizing these characteristics is crucial for enhancing diagnostic accuracy.

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.25-1524DOI Listing

Publication Analysis

Top Keywords

fbp method
12
myocardial perfusion
12
correction methods
12
coronary artery
12
created fbp
8
quantitative evaluation
8
ndb method
8
method
6
correction
5
[application method
4

Similar Publications

Background: We retrospectively examined image quality (IQ) of thin-slice virtual monochromatic imaging (VMI) of half-iodine-load, abdominopelvic, contrast-enhanced CT (CECT) by dual-energy CT (DECT) with deep learning image reconstruction (DLIR).

Methods: In 28 oncology patients with moderate-to-severe renal impairment undergoing half-iodine-load (300 mgI/kg) CECT by DECT during the nephrographic phase, we reconstructed VMI at 40-70 keV with a slice thickness of 0.625 mm using filtered back-projection (FBP), hybrid iterative reconstruction (HIR), and DLIR; measured contrast-noise ratio (CNR) of the liver, spleen, aorta, portal vein, and prostate/uterus; and determined the optimal keV to achieve the maximal CNR.

View Article and Find Full Text PDF

Systematic Review and Meta-Analysis of Radiation Dose Reduction Studies in Pediatric Head CT.

AJNR Am J Neuroradiol

March 2025

From the Department of Surgery, Division of Plastic and Reconstructive Surgery, Washington University School of Medicine, St. Louis, Missouri, USA (Kevin He, Adam Boukind, Anusha S. Sanka, Joseph G. Ribaudo, Sophia Chryssofos, Gary B. Skolnick, Kamlesh B. Patel), Bernard Becker Medical Library, Washington University School of Medicine, St. Louis, Missouri, USA (Lauren B. Yaeger), and the Mallinckrodt Institute of Radiology, Washington University School of Medicine, St. Louis, Missouri, USA (Allan M. Thomas, Ali Y. Mian).

Background: Conventional imaging protocols used in pediatric head CT scanning without specific adaptations to lower radiation dose or "standard dose" pediatric head CTs increase unnecessary radiation exposure. Modifying CT parameters, utilizing iterative reconstruction, and adopting specialized protocols are ongoing strategies to lower radiation dose in pediatric head CTs.

Purpose: This article will review studies reducing radiation exposure in pediatric patients undergoing head CT and provide metaanalysis of percent radiation dose reduction of the studies.

View Article and Find Full Text PDF

This study investigates Alzheimer's Disease (AD) dementia through [18F]Florbetapir ([18F]FBP) Positron Emission Tomography (PET) imaging. We employ Independent Component Analysis (ICA) to identify shared latent patterns across controls and individuals with Dementia. The dataset comprises PET brain images from 440 participants from the Alzheimer's Disease Neuroimaging Initiative (ADNI).

View Article and Find Full Text PDF

Background And Purpose: Dual-energy computed tomography (DECT) utilizes separate X-ray energy spectra to improve multi-material decomposition (MMD) for various diagnostic applications. However accurate decomposing more than two types of material remains challenging using conventional methods. Deep learning (DL) methods have shown promise to improve the MMD performance, but typical approaches of conducing DL-MMD in the image domain fail to fully utilize projection information or are under computationally inefficient iterative setup.

View Article and Find Full Text PDF

Positron emission tomography (PET) tracer binding may not be aligned with commonly used parcellations of neocortex [1]. Independent component analysis (ICA) can capture coactivated regions among participants that might serve as robust templates from a data-driven perspective. NeuroMark is a framework combining pre-defined templates with spatially constrained ICA, capturing a wide range of brain markers across imaging modalities [2],[3].