Background: Treatment planning systems (TPS) often exclude immobilization devices from optimization and calculation, potentially leading to inaccurate dose estimates. This study employed deep learning methods to automatically segment 3D-printed head and neck immobilization devices and evaluate their dosimetric impact in head and neck VMAT.

Methods: Computed tomography (CT) positioning images from 49 patients were used to train the Mask2Former model to segment 3D-printed headrests and MFIFs. Based on the results, four body structure sets were generated for each patient to evaluate the impact on dose distribution in volumetric modulated arc therapy (VMAT) plans: S (without immobilization devices), S (with MFIFs), S (with 3D-printed headrests), and S (with both). VMAT plans (P, P, P, and P) were created for each structure set. Dose-volume histogram (DVH) data and dose distribution of the four plans were compared to assess the impact of the 3D-printed headrests and MFIFs on target and normal tissue doses. Gafchromic EBT3 film measurements were used for patient-specific verification to validate dose calculation accuracy.

Results: The Mask2Former model achieved a mean average precision (mAP) of 0.898 and 0.895, with a Dice index of 0.956 and 0.939 for the 3D-printed headrest on the validation and test sets, respectively. For the MFIF, the Dice index was 0.980 and 0.981 on the validation and test sets, respectively. Compared to P, P reduced the V for PGTVnx, PGTVnd, PGTVrpn, PTV1, and PTV2 by 5.99%, 6.51%, 5.93%, 2.24%, and 1.86%, respectively(P ≤ 0.004). P reduced the same targets by 1.78%, 2.56%, 1.75%, 1.16%, and 1.48%(P < 0.001), with a 31.3% increase in skin dose (P < 0.001). P reduced the V by 9.15%, 10.18%, 9.16%, 3.36%, and 3.28% (P < 0.001), respectively, while increasing the skin dose by 31.6% (P < 0.001). EBT3 film measurements showed that the P dose distribution was more aligned with actual measurements, achieving a mean gamma pass rate of 92.14% under the 3%/3 mm criteria.

Conclusions: This study highlights the potential of Mask2Former in 3D-printed headrest and MFIF segmentation automation, providing a novel approach to enhance personalized radiation therapy plan accuracy. The attenuation effects of 3D-printed headrests and MFIFs reduce V and D for PTVs in head and neck cancer patients, while the buildup effects of 3D-printed headrests increases the skin dose (31.3%). Challenges such as segmentation inaccuracies for small targets and artifacts from metal fasteners in MFIFs highlight the need for model optimization and validation on larger, more diverse datasets.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11837297PMC
http://dx.doi.org/10.1186/s12885-025-13669-0DOI Listing

Publication Analysis

Top Keywords

immobilization devices
16
head neck
12
3d-printed headrests
12
segment 3d-printed
8
mask2former model
8
headrests mfifs
8
dose distribution
8
vmat plans
8
validation test
8
test sets
8

Similar Publications

Impact of a three-week full leg cast immobilization on infant crawling kinetics and spatiotemporal parameters.

PLoS One

March 2025

Wellstar College of Health and Human Services, Kennesaw State University, Kennesaw, Georgia, United States of America.

Despite its importance in the development of mobility in infants, there is a general lack of quantified data on infant crawling, and a specific lack of understanding of crawling kinetics, which includes the relative weightbearing and force generation among the four limbs. Moreover, because of the difficulty of measurement and study design, there is no longitudinal quantification of the impact of a perturbation to typical development, such as limb immobilization. This study measured kinetic and spatiotemporal outcomes in a typically developing infant prior to a fracture that necessitated a long-leg cast that immobilized the left knee and ankle, while crawling with the cast, and again one day after cast removal, and two weeks post.

View Article and Find Full Text PDF

A disposable, self-powered enzymatic biofuel cell (BFC) sensor integrated with a hollow microneedle array (HMNA) for glucose monitoring in interstitial fluid (ISF) is reported. The HMNA enables painless and minimally invasive ISF extraction. The BFC uses dehydrogenase (GDH) in conjunction with NAD, diaphorase (DI), and vitamin K (VK) serving as electron transfer mediators as the anode catalyst and Prussian blue (PB) as the electrochromic cathode.

View Article and Find Full Text PDF

Background And Purpose: Distal ulna fractures often occur in conjunction with distal radius fractures and other associated injuries. Currently, there are no satisfactory internal fixation systems available for addressing unstable distal ulna fractures, and a definitive consensus on the most effective treatment approach is still lacking. The objective of this research was to evaluate the clinical outcomes of using elastic stable intramedullary nails (ESIN) compared to locking compression plates (LCP) for treating unstable distal ulnar fractures in adults.

View Article and Find Full Text PDF

Objective: The primary aim of this investigation is to assess the effectiveness of implementing an innovative immobilization approach, spec ically the utilization of personalized open-face masks in combination with styrofoam fixation, for head and neck cancers receiving radiotherapy. The study seeks to evaluate the influence of this method on improving patients' precision in positioning and their overall comfort during the treatment process, in addition to exploring its potential capacity to mitigate the occurrence of anxiety and depression in this patient population.

Methods: A prospective, randomized controlled trial was undertaken to investigate the comparative efficacy of two immobilization approaches for the radiotherapy treatment of head and neck cancers.

View Article and Find Full Text PDF

Comparative efficacy of cast immobilization versus removable braces in patients with ankle fractures: a systematic review and meta-analysis.

BMC Musculoskelet Disord

March 2025

Department of Orthopedic Surgery, Hai'an People's Hospital, Zhongba Road 17, Hai'an, Nantong, Jiangsu, 226600, People's Republic of China.

Background: An effective and appropriate method to support the ankle joint optimally is particularly important during the healing phase of ankle fractures. The purpose of this review was to assess the functional outcomes, ankle-related quality of life, and associated complications of cast immobilization versus removable braces for the treatment of adult ankle fractures.

Methods: Studies comparing cast immobilization and removable braces after ankle fracture were included by systematic searches of PubMed/MEDLINE, Web of Science, Scopus, and EMBASE databases according to PRISMA (Preferred Reporting Items for Systematic Evaluation and Meta-Analysis Statements) guidelines.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!