Background: Malposition of the acetabular component during total hip arthroplasty (THA) is a leading cause of complications and need for revision. Robotic-assisted THA purports to improve accuracy of component positioning with many reports demonstrating over 92% of components within 10° of inclination and anteversion compared to intraoperative system output. This study aimed to evaluate the intraoperative accuracy output of acetabular cup position values using a handheld miniaturized portable navigation system (Naviswiss, AG) compared to postoperative computed tomography (CT) scans.

Methods: A total of 108 direct anterior approach THA surgeries using the intraoperative navigation device were performed over a 6-month period. Intraoperative device output for measured acetabular component inclination and anteversion were recorded and compared with values derived from postoperative CT scans.

Results: Postoperative CT analysis of acetabular component positioning demonstrated 97.22% and 94.44% were within 5° of intraoperative inclination and anteversion, respectively, compared to intraoperative values from the navigation unit. No CT demonstrated an absolute inclination or anteversion measurement difference more than 8° from the intraoperative navigation unit value. Overall, 92.59% of components were within 5° and 100% were within 8° for both inclination and anteversion compared to postoperative CT measurements.

Conclusions: This handheld portable navigation system yielded highly accurate intraoperative component positioning values confirmed by postoperative CT scans during direct anterior THA, possibly superior to reported robotic-assisted THA values. These smaller, portable, and more accessible intraoperative units may provide surgeons improved accuracy and availability in a number of surgical settings for use in THA.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11827077PMC
http://dx.doi.org/10.1016/j.artd.2024.101511DOI Listing

Publication Analysis

Top Keywords

inclination anteversion
20
acetabular component
16
component positioning
16
anteversion compared
12
intraoperative
9
handheld portable
8
robotic-assisted tha
8
compared intraoperative
8
portable navigation
8
navigation system
8

Similar Publications

Background: Prosthetic impingement after total hip arthroplasty (THA) has been associated with instability and may be a cause of accelerated polyethylene wear and pain. Previous retrieval studies report a high prevalence of impingement in acetabular liners. Robotic technology has the potential to reduce THA instability as it enables technical precision and optimizes implant positioning.

View Article and Find Full Text PDF

Background: The present study aimed to examine the impact of prior experience of computed tomography-based navigated THA (CTN-THA) on the accuracy of cup placement in robotic arm-assisted total hip arthroplasty via direct anterior approach using the Mako system (Mako-THA).

Methods: We evaluated the first 60 hips that underwent Mako-THA performed by surgeons with and without CTN-THA experience and compared the absolute error of pre-operative and post-operative cup placement using a three-dimensional template between the two groups.

Results: The CTN-user group demonstrated significantly better results in radiographic inclination, y-(sagittal) axis placement, and z-(vertical) axis placement than the non-CTN-user group (CTN-user group: 1.

View Article and Find Full Text PDF

Purpose: Accurate cup placement is challenging in total hip arthroplasty (THA) for dysplastic hip osteoarthritis (DHOA) because of the complex morphology of the acetabulum. Studies have reported good accuracy for total hip arthroplasty (THA) using computed tomography-based navigation (CTN); however, in recent years, portable navigation (PN) has become more widely applied because of its low cost and ease of use. This study aimed to compare the accuracy of portable navigation with that of CT-based navigation.

View Article and Find Full Text PDF

Background: Patients often report leg-length discrepancy (LLD) after total hip arthroplasty (THA). The factors behind self-reported LLD (sLLD) are unclear. Robot-assisted THA (RA-THA) offers improved precision, but its effect on sLLD has not been well studied.

View Article and Find Full Text PDF

Background: Computer-assisted surgical navigation was designed to reduce the rates of acetabular component malpositioning, a common cause for revision following primary total hip arthroplasty (THA). We compared intraoperative measures such as radiation exposure, radiographic acetabular cup placement, and outcomes of patients who underwent a THA using computer-assisted surgical navigation with augmented reality (AR) compared to those who underwent traditional manual THA with fluoroscopic assistance (FA).

Methods: We retrospectively reviewed 115 hips who underwent primary direct anterior THAs using the intraoperative AR surgical navigation system between 3/30/2021 and 12/30/2022.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!