Introduction: Hemolysis is a known interference factor that has been found to show erroneous effect. Present study analyzes the impact of hemolysis on the concentrations of protein biomarkers of Alzheimer's disease (Aβ42, t-Tau, p-Tau181) along with novel proteins which are currently under investigation (SIRT1,SIRT2,SIRT6,FOXO3A, NFL, Aβ40, GFAP).
Methods: Plasma samples were grouped into two categories: hemolyzed and non-hemolyzed groups. Degree of hemolysis (in percentage) was separately analyzed using Single molecule array (SIMOA) technology. Quantitative analysis for hemolyzed and non-hemolyzed samples were done using surface plasmon resonance (SPR) technology.
Results: The SIMOA analysis indicated that at high levels of hemolysis (1000 mg/dL) there was an increase in NFL protein level up to approximately 30 % whereas p-Tau181 did not show much interference even at higher hemolysate concentration. Aβ40, Aβ42 and GFAP showed modest effect up to hemolysis of 250mg/dL-500 mg/dL. SPR analysis of total Tau (t-Tau), p-Tau181, SIRT1, SIRT6 showed the consistency in the result and there was no significant difference in hemolyzed plasma compared to non-hemolyzed samples. Aβ42 and FOXO3A showed decline in hemolyzed plasma compared to non-hemolyzed samples (4.34 ± 0.18ng/ul; 4.95 ± 0.19ng/ul) and (3.83 ± 0.34ng/ul; 5.12 ± 0.46ng/ul), respectively whereas, a significant increase in the concentration was observed for SIRT2; 2.4 ± 0.10ng/ul in hemolyzed compared to 1.30 ± 0.22ng/ul in non-hemolyzed group.
Conclusions: High grade hemolysis leads to altered protein concentration associated with neurodegeneration. Present study emphasizes the need to have pre-analytical inspection for hemolysis detection especially in a multicentric biomarker study.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11791351 | PMC |
http://dx.doi.org/10.1016/j.plabm.2025.e00455 | DOI Listing |
Pract Lab Med
April 2025
Venu Geriatric Care Center, Venu Eye Institute, New Delhi, India.
Introduction: Hemolysis is a known interference factor that has been found to show erroneous effect. Present study analyzes the impact of hemolysis on the concentrations of protein biomarkers of Alzheimer's disease (Aβ42, t-Tau, p-Tau181) along with novel proteins which are currently under investigation (SIRT1,SIRT2,SIRT6,FOXO3A, NFL, Aβ40, GFAP).
Methods: Plasma samples were grouped into two categories: hemolyzed and non-hemolyzed groups.
The Working Group Preanalytical Phase of the European Federation of Clinical Chemistry and Laboratory Medicine advises suppressing haemolysis-sensitive tests when haemolysis is clinically significant, as improper specimen handling can rupture red blood cells, increasing potassium levels. Thus, a correctly repeated blood sample should show lower potassium levels than a haemolysed one. This study aimed to determine the prevalence of haemolysed samples with potassium levels below the reference range and whether this predicts hypokalemia in repeated collections.
View Article and Find Full Text PDFAnimals (Basel)
June 2024
Veterinary Faculty, University of Ljubljana, Gerbičeva 60, 1000 Ljubljana, Slovenia.
Hemolysis is a common cause of errors in laboratory tests as it affects blood parameters and leads to a positive or negative bias. This study aims to examine the relationship between the level of hemolysis (expressed as cell-free hemoglobin concentration, g/L) and the variability of metabolic and endocrine parameters and to determine the threshold level of hemolysis that causes an analytically and clinically significant bias for the twenty most frequently examined blood parameters in cows. Paired blood samples of 10 mL each were obtained from 30 cows.
View Article and Find Full Text PDFClin Chem Lab Med
September 2024
Department of Emergency Medicine, Baylor College of Medicine, Houston, TX, USA.
Objectives: The destruction of red cells during blood collection or with the processing of the sample continues to occur at a high rate, especially among emergency department (ED) patients. This can produce pre-analytical laboratory errors, particularly for potassium. We determined the incidence of hemolyzed samples and discuss the potential medical impact for hypokalemic patients who potassium level is artificially normal (pseudoeukalemia).
View Article and Find Full Text PDFClin Chem Lab Med
January 2024
Department of Laboratory Medicine, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
Objectives: Several studies have demonstrated that synovial calprotectin is a highly accurate biomarker in diagnosing periprosthetic joint infections (PJI). Assuring reliability is of great importance and coincides with adequate preanalytical handling. This study focuses on potentially interfering factors.