Background: Management of respiratory motion during radiation therapy is essential for accurate dose delivery and minimizing the risk to organs. In diagnostic imaging, respiratory monitoring is required for confirmation of breath-hold and four-dimensional computed tomography (CT) reconstruction. Although respiratory monitoring systems are widely used in radiation therapy, they are not often used for diagnostic imaging, where they could improve image quality.
Purpose: The purpose of this study was to use a millimeter-wave sensor (MWS) to noninvasively visualize respiratory motion, confirm breath-holding, and explore the potential for clinical implementation of an MWS in diagnostic x-ray imaging, CT, and radiation therapy.
Methods: A 24 GHz microMWS was used in this study. The MWS directionality was determined using a radio-wave dark-box system. An antenna directionality test evaluated the effective azimuthal and elevational beamwidths. Respiratory waveforms were detected by optimizing the fast Fourier transform threshold and the cutoff frequencies of the bandpass filter. To confirm the reproducibility of the MWS, the detected waveforms were compared with those of a respiratory motion phantom (QUASAR), the amplitude of motion of which could be controlled. The time from valley to peak of the waveforms obtained by normalized MWS and the QUASAR were compared. The MWS was used to acquire respiratory waveforms of 20 healthy volunteers (including an infant and a child) in geometries adopted during chest CT (supine position; anteroposterior view; source-to-surface distance, 400 mm) and chest x-ray imaging (standing position; posteroanterior view; source-to-surface distance, 1800 mm).
Results: The effective azimuthal and elevational beamwidths of the MWS were approximately ± 20° and ± 40°, respectively. By optimizing the acquisition parameters (high-sensitivity setting; with noise cancelling; frequency range, 10-20 min), the waveforms detected using the MWS approximately matched those of the respiratory motion phantom at all amplitudes. The MWS was also used to confirm breath-holding in 18 volunteers in both supine (anteroposterior view) and standing (posteroanterior view) positions. In addition, for an infant and a child who were unable to follow the instruction to stop breathing, a visual count of their inhalations matched the number of respiratory cycles measured using the MWS.
Conclusion: The 24 GHz MWS successfully monitored respiratory motion and breath-holding during radiographic and CT imaging. With effective directionality and stability, this system holds promise for clinical management of respiratory motion during diagnostic imaging and radiation therapy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.17616 | DOI Listing |
Cureus
December 2024
Department of Rehabilitation Medicine, School of Medicine, Showa University, Tokyo, JPN.
Tetanus is a rare but life-threatening neurological disorder caused by neurotoxins produced by . Although mortality rates have significantly decreased with modern intensive care, severe cases remain challenging due to prolonged Intensive Care Unit (ICU) stays, complications, and rehabilitation barriers. We report the case of an 81-year-old male with a history of hypertension and femoral neck fracture who developed severe tetanus following a contaminated forehead laceration.
View Article and Find Full Text PDFJ Exp Biol
January 2025
African Robotics Unit, University of Cape Town, Cape Town, 7700, Western Cape, South Africa.
Understanding and monitoring wildlife behavior is crucial in ecology and biomechanics, yet challenging due to the limitations of current methods. To address this issue, we introduce WildPose, a novel long-range motion capture system specifically tailored for free-ranging wildlife observation. This system combines an electronically controllable zoom-lens camera with a LiDAR to capture both 2D videos and 3D point cloud data, thereby allowing researchers to observe high-fidelity animal morphometrics, behavior and interactions in a completely remote manner.
View Article and Find Full Text PDFMed Phys
January 2025
Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osakasayama, Osaka, Japan.
Background: Management of respiratory motion during radiation therapy is essential for accurate dose delivery and minimizing the risk to organs. In diagnostic imaging, respiratory monitoring is required for confirmation of breath-hold and four-dimensional computed tomography (CT) reconstruction. Although respiratory monitoring systems are widely used in radiation therapy, they are not often used for diagnostic imaging, where they could improve image quality.
View Article and Find Full Text PDFExp Physiol
January 2025
Neurovascular Research Laboratory, Faculty of Life Sciences and Education, University of South Wales, Pontypridd, UK.
Fenestrated/branched endovascular aortic repair emerges as the primary therapeutic modality for intricate aortic pathologies encompassing the paravisceral and thoracoabdominal segments, where bridging stent grafts (BSGs) play a vital role in linking the primary aortic endograft with target vessels. Bridging stent grafts can be categorized mainly into self-expanding stent grafts (SESGs) and balloon-expandable stent grafts (BESGs). Physiological factors significantly influence post-complex endovascular aortic repair BSG behaviour, impacting clinical outcomes of SESGs and BESGs in different but overlapping ways.
View Article and Find Full Text PDFMucus plays an integral role for the barrier function of many epithelial tissues. In the human airways, mucus is constantly secreted to capture inhaled microbes and pollutants and cleared away through concerted ciliary motion. Many important respiratory diseases exhibit altered mucus flowability and impaired clearance, contributing to respiratory distress and increased risk of infections.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!