Introduction: Kinematic alignment (KA) in total knee arthroplasty (TKA) is by definition a pure femoral resurfacing procedure aiming to restore the individual prearthritic anatomy. However, when a 2 mm compensation is systematically used on the worn side, the variability in cartilage thickness in the unworn compartment might alter the accuracy of the technique. This study aimed to validate two intraoperative femoral cartilage thickness measurement techniques by comparing them to the photographic method, which measures cartilage thickness through pixel analysis of bone-cut images. The study hypothesized that the two intraoperative methods are comparable and similarly accurate within 0.5 mm of the photographic method.

Methods: Seventy cartilage thickness measurements from seventy patients with end-stage knee osteoarthritis were prospectively collected. Two intraoperative techniques were evaluated: the electrocautery tip method (Method A) and the ruler method (Method B), performed before and after distal femoral bone resections, respectively. The postoperative photographic analysis (Method C) served as the reference method. Measurements were rounded to the nearest 0.5 mm for consistency. Data were analyzed using Kruskal-Wallis test, Wilcoxon rank-sum tests, Spearman's rank correlation, percentage of agreement and intraclass correlation coefficients (ICCs).

Results: No significant differences were observed between Method A and Method B in measuring femoral cartilage thickness. Agreement with Method C was 100% for Method B and 85% for Method A. In the 15% of discordant cases, Method A overestimated the measurements by one category of 0.5 mm compared to Method C. Correlation coefficients between the methods were high ( = 0.88-1.0). Intra- and interobserver reliability was high for all methods (ICCs 0.91-0.95).

Discussion: Both intraoperative methods are reliable and comparable to the photographic method when rounded to the closest 0.5 mm, with no significant differences among them. The electrocautery method has the added advantage of measuring cartilage thickness before bone cuts are performed.

Level Of Evidence: Level IV.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11763056PMC
http://dx.doi.org/10.1002/jeo2.70155DOI Listing

Publication Analysis

Top Keywords

cartilage thickness
28
method
16
method method
12
total knee
8
knee arthroplasty
8
kinematic alignment
8
femoral cartilage
8
photographic method
8
intraoperative methods
8
electrocautery method
8

Similar Publications

Deep Learning Superresolution for Simultaneous Multislice Parallel Imaging-Accelerated Knee MRI Using Arthroscopy Validation.

Radiology

January 2025

From the Department of Radiology, Division of Musculoskeletal Radiology, NYU Grossman School of Medicine, 660 1st Ave, 3rd Fl, Rm 313, New York, NY 10016 (S.S.W., J.V., R.K., E.H.P., J.F.); Department for Diagnostic and Interventional Radiology, Eberhard Karls University Tübingen, University Hospital Tübingen, Tübingen, Germany (S.S.W.); Department of Radiology, University Hospital Basel, Basel, Switzerland (J.V.); Department of Radiology, Hospital do Coraçao, São Paulo, Brazil (T.C.R.); Academic Surgical Unit, South West London Elective Orthopaedic Centre (SWLEOC), London, United Kingdom (D.D.); Department of Radiology, Balgrist University Hospital, Zurich, Switzerland (B.F.); Department of Radiology, Jeonbuk National University Hospital, Jeonju, Republic of Korea (E.H.P.); Research Institute of Clinical Medicine of Jeonbuk National University, Biomedical Research Institute of Jeonbuk National University Hospital, Jeonju, Republic of Korea (E.H.P.); Medscanlagos Radiology, Cabo Frio, Brazil (A.S.); Centre for Data Analytics, Bond University, Gold Coast, Australia (S.E.S.); Siemens Healthineers AG, Erlangen, Germany (I.B.); and Siemens Medical Solutions USA, Malvern, Pa (G.K.).

Background Deep learning (DL) methods can improve accelerated MRI but require validation against an independent reference standard to ensure robustness and accuracy. Purpose To validate the diagnostic performance of twofold-simultaneous-multislice (SMSx2) twofold-parallel-imaging (PIx2)-accelerated DL superresolution MRI in the knee against conventional SMSx2-PIx2-accelerated MRI using arthroscopy as the reference standard. Materials and Methods Adults with painful knee conditions were prospectively enrolled from December 2021 to October 2022.

View Article and Find Full Text PDF

Secondary Healing Over NovoSorb BTM in the Management of Complex Nasal Defects.

Plast Reconstr Surg Glob Open

January 2025

From the Department of Plastic and Reconstructive Surgery, Beaumont Hospital, Dublin, Ireland.

Background: The use of PolyNovo NovoSorb biodegradable temporizing matrix (BTM) has become popular for the management of complex wounds, often with the subsequent application of a split-thickness skin graft. To date, there are no published case series describing healing via secondary intention over BTM in the management of small, complex nasal wounds following excision of skin malignancy in the elective setting.

Methods: This study consisted of a single-center, consecutive case series.

View Article and Find Full Text PDF

Objective: To determine whether and when early division of the paramedian forehead flap may be safely performed.

Data Sources: PubMed (NLM), Scopus (Elsevier), and Embase (Elsevier).

Methods: A systematic search of PubMed, Scopus, and Embase was conducted according to PRISMA guidelines.

View Article and Find Full Text PDF

Zonal Characteristics of Collagen Ultrastructure and Responses to Mechanical Loading in Articular Cartilage.

Acta Biomater

January 2025

Biomedical Engineering, College of Engineering, Mathematics and Physical Sciences, University of Exeter, UK. Electronic address:

The biomechanical properties of articular cartilage arise from a complex bioenvironment comprising hierarchically organised collagen networks within the extracellular matrix (ECM) that interact with the proteoglycan-rich interstitial fluid. This network features a depth-dependent fibril organisation across different zones. Understanding how collagen fibrils respond to external loading is key to elucidating the mechanisms behind lesion and managing degenerative conditions like osteoarthritis.

View Article and Find Full Text PDF

Introduction: Kinematic alignment (KA) in total knee arthroplasty (TKA) is by definition a pure femoral resurfacing procedure aiming to restore the individual prearthritic anatomy. However, when a 2 mm compensation is systematically used on the worn side, the variability in cartilage thickness in the unworn compartment might alter the accuracy of the technique. This study aimed to validate two intraoperative femoral cartilage thickness measurement techniques by comparing them to the photographic method, which measures cartilage thickness through pixel analysis of bone-cut images.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!