Impact of a new image enhancement technology on the nuchal translucency thickness.

Arch Gynecol Obstet

Department of Obstetrics and Gynaecology, University Hospital Tübingen, University of Tübingen, Calwerstrasse 7, 72076, Tübingen, Germany.

Published: January 2025

Objective: To examine the impact of a new image enhancement technique on the distribution of NT measurements.

Methods: In this retrospective study, nuchal translucency (NT) images that were taken with the GE Voluson E22 ultrasound machine between May and September 2024 were collected. One operator took manual and automated NT measurements in an NT image without radiant enhancement mode followed by automated measurements in NT images with minimum, medium and maximum radiant mode. The automated measurement in the NT images without radiant mode were considered gold standard. The relative bias of the gold standard, the manual measurement and the automated measurements with the radiant enhancement technology were compared using median and 25-75th interquartile range as well as by the Wilcoxon test.

Results: The database search yielded 352 NT images from 101 pregnant women. Average maternal age of the study population was 34.0 years and average crown-rump length was 68.3. The median NT thickness was 1.7 mm for both the manual and automated measurements done without the use of radiant mode. The median automated measurement with minimum, medium and maximum radiant mode was 2.00, 2.05 and 2.10 mm, respectively. All automated measurements were significantly higher than the gold standard.

Discussion: The use of the radiant mode results in an increase in the NT thickness. This effect should be taken into account when calculating the risk for chromosomal abnormalities.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00404-024-07924-1DOI Listing

Publication Analysis

Top Keywords

automated measurements
20
radiant mode
20
impact image
8
image enhancement
8
enhancement technology
8
nuchal translucency
8
manual automated
8
radiant enhancement
8
mode automated
8
minimum medium
8

Similar Publications

Background: This special section underscores the potential of multimodal measurement approaches to transform psychotherapy research. A multimodal approach provides a more comprehensive understanding than any single modality (type of collected information) can provide on its own.

Methods: Traditionally, clinicians and researchers have relied on their intuition, experience, and training to integrate different types of information in a psychotherapy session/treatment.

View Article and Find Full Text PDF

Purpose: To develop an artificial intelligence (AI) algorithm for automated measurements of spinopelvic parameters on lateral radiographs and compare its performance to multiple experienced radiologists and surgeons.

Methods: On lateral full-spine radiographs of 295 consecutive patients, a two-staged region-based convolutional neural network (R-CNN) was trained to detect anatomical landmarks and calculate thoracic kyphosis (TK), lumbar lordosis (LL), sacral slope (SS), and sagittal vertical axis (SVA). Performance was evaluated on 65 radiographs not used for training, which were measured independently by 6 readers (3 radiologists, 3 surgeons), and the median per measurement was set as the reference standard.

View Article and Find Full Text PDF

Microtiter-plate-based systems are unified platforms of high-throughput experimentation (HTE). These polymeric devices are used worldwide on a daily basis-mainly in the pharmaceutical industry-for parallel syntheses, reaction optimization, various preclinical studies and high-throughput screening methods. Accordingly, laboratory automation today aims to handle these commercially available multiwell plates, making developments focused on their modifications a priority area of modern applied research.

View Article and Find Full Text PDF

Validation of an automated quality control method to test sterility of two advanced therapy medicinal products: Mesenchymal stromal cells and their extracellular vesicles.

Hematol Transfus Cell Ther

November 2024

Hospital São Rafael, Salvador, Bahia, Brazil; Instituto D'Or de Pesquisa e Ensino (IDOR), Salvador, Bahia, Brazil; Instituto Gonçalo Moniz, FIOCRUZ, Salvador, Bahia, Brazil. Electronic address:

Mesenchymal stromal cells are multipotent cells present in various tissues that are widely studied for relevant therapeutic potential due to their paracrine immunomodulatory and tissue regenerating properties. Many mesenchymal stromal cell-based products are under investigation for the treatment of different clinical conditions. Recently, the therapeutic potential of the extracellular vesicles released by these cells has been under focus, with emphasis on clinical translation.

View Article and Find Full Text PDF

Accurate urine sodium measurements at home using Point of Care Testing in patients with Short Bowel Syndrome.

Clin Nutr ESPEN

January 2025

Department of Gastroenterology and Hepatology, Intestinal Failure Unit, Radboud University Medical Centre Nijmegen, Geert Grooteplein 10, 6500 HB, Nijmegen, The Netherlands. Electronic address:

Background And Aims: Measurement of the urine sodium concentration (USC) is a simple procedure that in many patients adequately indicates their hydration status. This is of particular importance in patients suffering from short bowel syndrome (SBS), who may very rapidly dehydrate and are at risk for permanently compromising their kidney function. A point of care test (POCT) that allows reliable measurement of USC would enable these patients to effectively evaluate their sodium- and water balance in the at home setting, thereby avoiding hospital visits and delayed test results.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!