The use of an alteplase (Activase) purge solution to address Impella ventricular assist device "thrombosis" or "purge system occlusion" has been mainly documented with earlier generation Impella devices (CP, 2.5, 5.0). Here, we report the use of an alteplase purge solution to manage Impella 5.5 purge system occlusion in a 31-year-old male admitted to the cardiac care unit in cardiogenic shock and listed for a heart transplant. Throughout the purge system occlusion, the patient demonstrated hemodynamic stability and overall pump flow and cardiac output were preserved. Initially, there was a lack of response in purge flow and pressure observed at the lower concentration of alteplase purge solution (0.04 mg/ml), yet after using the alteplase 4 mg/50 ml purge (0.08 mg/ml) solution concentration, a response was seen. No bleeding or hemodynamic complications were observed. In addition, a suggested management workflow and review of the case reports and case series published to date regarding Impella purge system occlusion is included in this article.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAT.0000000000002378DOI Listing

Publication Analysis

Top Keywords

purge solution
16
purge system
16
system occlusion
16
alteplase purge
12
purge
9
impella ventricular
8
ventricular assist
8
assist device
8
impella purge
8
alteplase
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!