Background: Direct oral anticoagulants (DOACs) can interfere with coagulation analyses, causing erroneous results such as false-positive lupus anticoagulant and false-normal antithrombin, threatening patient safety when overlooked. A test using a prothrombin time quotient method to detect DOAC presence in plasma samples is now commercially available, the MRX PT DOAC, with the result expressed as Clot Time Ratio (CTR).
Objectives: Evaluate the ability of MRX PT DOAC to identify interfering apixaban or rivaroxaban concentrations, identify non-interfering or interfering patient samples, and detect whether a patient is on DOAC treatment.
Methods: An MRX PT DOAC reference interval was established, and MRX PT DOAC results related to FXa inhibitor concentrations. MRX PT DOAC patient plasma results were related to indications of analytical interference with lupus anticoagulant and antithrombin analyses and with the patients' anticoagulant drug use, according to medical records.
Results: The MRX PT DOAC CTR reference interval was 0.98-1.38. MRX PT DOAC apixaban and rivaroxaban interference detection sensitivity was 1.00 (0.96 - 1.0). For subgroups from 315 patient plasma samples, the negative predictive values estimates were 0.84 or above (95% confidence interval minimum 0.64) for excluding analysis interference with the lupus anticoagulant and antithrombin analyses and for excluding DOAC treatment as per medical records. All 9 interference-indicating DOAC samples were identified for the lupus anticoagulant analyses, 5 out of 11 for antithrombin.
Conclusions: The MRX PT DOAC is a potential screening tool for interference. Confirming results in a larger DOAC patient group and further investigating the antithrombin interference detection uncertainty is important.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1080/00365513.2025.2457115 | DOI Listing |
Scand J Clin Lab Invest
January 2025
Department of Biomedical and Clinical Sciences, Linköping University, Linköping, Sweden.
Background: Direct oral anticoagulants (DOACs) can interfere with coagulation analyses, causing erroneous results such as false-positive lupus anticoagulant and false-normal antithrombin, threatening patient safety when overlooked. A test using a prothrombin time quotient method to detect DOAC presence in plasma samples is now commercially available, the MRX PT DOAC, with the result expressed as Clot Time Ratio (CTR).
Objectives: Evaluate the ability of MRX PT DOAC to identify interfering apixaban or rivaroxaban concentrations, identify non-interfering or interfering patient samples, and detect whether a patient is on DOAC treatment.
J Thromb Haemost
December 2024
Department of Medicine, McMaster University, Hamilton, Ontario, Canada.
Background: Although routine monitoring is not needed for direct oral anticoagulants (DOACs), knowing if a clinically relevant DOAC level is present can be critical, especially in cases of severe bleeding or urgent surgery. Rapid assays to exclude these levels are necessary but not widely available.
Objectives: To determine the test performance of MRX PT DOAC for excluding clinically relevant DOAC drug levels.
Sci Rep
March 2024
Department of Vascular Diseases, University Medical Centre Ljubljana, Ljubljana, Slovenia.
There are clinical situations where information about the anticoagulant effects of Apixaban could be useful. Specialised methods for measuring Apixaban concentrations are not available at all medical laboratories while methods for measuring the functional effects of Apixaban, using clot time ratio (CTR), can be performed in most medical laboratories around the clock using well-established measurement procedures. The aim of this study was to investigate CTR in trough and peak samples during Apixaban treatment of atrial fibrillation and to correlate the findings to bleeds and thrombotic events.
View Article and Find Full Text PDFThromb J
March 2024
Department of Vascular Diseases, University Medical Centre Ljubljana, Ljubljana, Slovenia.
Background: There are situations where information about the anticoagulant effects of Rivaroxaban could be clinically useful. Methods for measuring Rivaroxaban concentrations are not available at all medical laboratories while the test MRX PT DOAC for measuring the functional effects of Rivaroxaban, in CTR (Clot Time Ratio), can be made available around the clock. The objectives of this study were to investigate CTR in trough and peak samples during Rivaroxaban treatment of atrial fibrillation and to correlate the findings to bleeding episodes.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!