Background: Continuous spinal anesthesia (CSA) offers precise, prolonged neuraxial anesthesia suitable for high-risk patients. This technique minimizes hemodynamic instability but comes with notable challenges. Vigilant catheter management is crucial to avoid complications, including the risk of catheter retention. We report a case where the mechanical properties of a spinal catheter led to diagnostic uncertainty.

Case: . Following successful insertion, the surgery proceeded uneventfully. During catheter removal, the expected tip markings were absent, raising concern for a retained fragment. Imaging confirmed no retained foreign body. To investigate further, stretching of a new catheter replicated the disappearance of tip markings. This finding suggested that the catheter's material had the potential to stretch under tension, leading to fading of the markings and mimicking a retained fragment.

Conclusions: This case underscores the need for awareness of the mechanical properties of spinal catheters, especially their potential for stretching and loss of tip markings. Understanding these characteristics can prevent unnecessary imaging and reduce patient and clinician anxiety.

Download full-text PDF

Source
http://dx.doi.org/10.1136/rapm-2024-106235DOI Listing

Publication Analysis

Top Keywords

spinal catheter
8
mechanical properties
8
properties spinal
8
catheter
7
disappearance spinal
4
catheter catheter
4
catheter stretching
4
stretching case
4
case report
4
report background
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!