Background: This study aimed to quantify radiation doses during navigational bronchoscopy procedures, comparing them with reported cohorts and evaluating the LungVision (Body Vision Medical Inc.) system's efficacy in dose reduction.
Methods: This retrospective observational study included 52 consecutive navigational bronchoscopy cases, categorized into 4 imaging groups based on the C-arm: Cios Spin (Siemens Healthineers), or OEC 9900 (GE HealthCare); and the 3D tomographic imaging algorithm: Cios Spin's onboard imaging, or LungVision's AI-driven imaging. Patient and lesion data, outcomes, and radiation indices were collected. Existing literature on 3D image guidance for bronchoscopic lung nodules was reviewed to compare reported radiation doses.
Results: Combining LungVision with Cios Spin significantly reduced radiation dose in all cases compared with using Cios Spin alone: Cumulative air kerma (Ka,r) reduced from 238.7 to 119.1 mGy (P=0.03), and air kerma-area product (KAP) decreased from 28.19 to 15.09 Gy·cm2 (P=0.03). For biopsy cases, LungVision led to notable dose reductions: Ka,r of 279 to 129.1 mGy, and KAP of 30.70 to 16.27 Gy·cm2. LungVision notably reduced radiation indices in 7 paired spins, isolating the 3D imaging algorithm as the sole variable with the same Cios Spin C-arm. A literature review provides additional context on radiation for bronchoscopic biopsies.
Conclusion: Following the "as low as reasonably achievable" (ALARA) principle minimizes ionizing radiation exposure, benefiting patients and operators. Physicians should compare baseline radiation levels with the literature and adopt dose-reduction techniques. LungVision's lower dose indices render it effective for real-time 3D imaging during navigational bronchoscopy while reducing radiation dose.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/LBR.0000000000001001 | DOI Listing |
Int J Surg
January 2025
Carcinoma Department of Traditional Chinese Medicine, Dianjiang People's Hospital of Chongqing, Chongqing, PR China.
The widespread adoption of high-resolution computed tomography (CT) screening has led to increased detection of small pulmonary nodules, necessitating accurate localization techniques for surgical resection. This review examines the evolution, efficacy, and safety of various localization methods for small pulmonary nodules. Studies focusing on localization techniques for pulmonary nodules ≤30 mm in diameter were included, with emphasis on technical success rates and complication profiles.
View Article and Find Full Text PDFDiseases
January 2025
Department of Pneumonology, Medical School, Democritus University of Thrace, University General Hospital Dragana, 68100 Alexandroupolis, Greece.
: Eosinophilic pneumonias denote a rare condition, wherein infiltrating eosinophilic granulocytes accumulate within the lung parenchyma. Although eosinophilic pneumonias may be idiopathic, they are also associated with secondary causes. More than 110 medications have been linked to eosinophilic pneumonia, including several antidepressants.
View Article and Find Full Text PDFCureus
December 2024
Department of Pulmonary Medicine, All India Institute of Medical Sciences, Jodhpur, Jodhpur, IND.
Background Airway foreign body aspiration is an emergency predominantly observed in children and the elderly. However, it also occurs in adults, presenting with a variety of symptoms. Both rigid and flexible bronchoscopies are employed for foreign body retrieval.
View Article and Find Full Text PDFJ Bronchology Interv Pulmonol
April 2025
Thoracic Surgery, BASS Medical Group, Walnut Creek, CA.
Background: This study aimed to quantify radiation doses during navigational bronchoscopy procedures, comparing them with reported cohorts and evaluating the LungVision (Body Vision Medical Inc.) system's efficacy in dose reduction.
Methods: This retrospective observational study included 52 consecutive navigational bronchoscopy cases, categorized into 4 imaging groups based on the C-arm: Cios Spin (Siemens Healthineers), or OEC 9900 (GE HealthCare); and the 3D tomographic imaging algorithm: Cios Spin's onboard imaging, or LungVision's AI-driven imaging.
Best Pract Res Clin Anaesthesiol
March 2024
1400 Holcombe Blvd, FC 13.2000, Houston, TX, 77030, USA. Electronic address:
Lung cancer is among one of the most commonly diagnosed malignancies and is the leading cause of cancer-related mortality in both men and women globally, with an estimated 1.8 million deaths annually. Moreover, it is also the leading cause of cancer related deaths in the United States (U.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!