The present study describes the "fit for purpose" testing and the independent product-specific GMP validation of the monocyte activation test (MAT) to detect pyrogenic and pro-inflammatory contaminants, MAT Method A, Quantitative Test (European Pharmacopoeia, Ph. Eur. chapter 2.6.30, 2017). A fit for purpose study was carried out to ensure that the chosen MAT set-up (cryopreserved PBMC, IL-6 detection) can reliably discriminate between batches of product containing pyrogenic contaminants below the contaminants limit concentration, CLC, from batches containing pyrogenic contaminants above the CLC. Such testing is carried out once, before the chosen MAT set-up is used for subsequent product testing to show that the incidence of false positives (pyrogen-negative (
Download full-text PDF
Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11721511 PMC http://dx.doi.org/10.1016/j.crtox.2024.100206 DOI Listing Publication Analysis